14
ENGLISH
PRODUCT SPECIFICATIONS
Copy of the EC Declaration of conformity can be requested to Gima S.p.A. - Via Marconi, 1 - 20060 –
Gessate, Milan (Italy)
MMAD = Mass Median Aero-
dynamic Diameter
GSD = Geometric Standard
Deviation
NB: The measures and the curves are not valid in the event of medications supplied in high viscosity
suspension.
TYPE (Directive 93/42/EEC)
Medical device Class IIa
MODEL
Eolo
POWER SUPPLY
230V~ / 50Hz
ABSORBED POWER
170 VA
FUSE
F 1 x 1.6A L 250V
MAXIMUM PRESSURE
250 kPa (2.5 Bar)
MAXIMUM FLOW (in the compressor)
14 L/min
OPERATING PRESSURE
110 kPa (1.10 Bar)
OPERATIONAL FLOW
5.0 L/min at 110 kPa
ATOMIZATION
0.40 ml/min (with 4ml solution NaCl 0.9%)
MMAD (measured in accordance
with EN 13544-1)
2.44
GSD
2.87
WEIGHT
1.65 Kg
SIZE
148 (L) x 223 (H) x 124 (P) mm
MAXIMUM NOISE LEVEL
55 dB (A)
FUNCTIONING
Continuous
MINIMUM NEBULIZER VOLUME LEVEL
2ml
MAXIMUM NEBULIZER VOLUME LEVEL 6ml
OPERATING CONDITIONS
Room temperature: 5 ÷ 40°C
Percentage of humidity in the environment: 10 ÷ 93% RH
Atmospheric pressure: 700 ÷ 1060 hPa
STORAGE AND TRANSPORT
CONDITIONS
Room temperature: -25 ÷ 70°C
Percentage of humidity in the environment: 0 ÷ 93% RH
Atmospheric pressure: 500 ÷ 1060 hPa
Summary of Contents for 28091
Page 64: ...64 Eolo 230V 50Hz GIMA 1 2 3 4 Gima S p A 5 Gima...
Page 65: ...65 6 7 Eolo 8 9 36 10 11 ISO 10993 1 12 EN 60601 1 13 14 Gima S p A 15 16...
Page 67: ...67 MMAD GSD Eolo F 1 6A L 250V...
Page 68: ...68 Gima GIMA 1 2 3 1 5 40 C 2 60 40 40 C 1 2 3 4 5 Gima 1 2 3 4 5 Gima...
Page 69: ...69 H HI FLO HI FLO 1 2 3 4 5 6 7 8 6 120 25 7 6 8 5 1 3 4 2...
Page 70: ...70 1500 5 5 Gima 2 5 4 6 1 I O 5 4 6 1 2 3...
Page 75: ...75 GIMA 12 GIMA GIMA software GIMA...
Page 76: ......