USE

The 

28085

 ultrasound aerosol is used to treat asthma, chronic obstructive pulmonary disorders and other 

breathing problems in which a therapy with aerosol is required. It is suitable for use in one single patient.

CLEANING AND ORDINARY MAINTENANCE

It is important to clean the Aerosol before use and after every use, drying it well in every part.

Remember: remove the plug and do not plunge in water

1. Remove the cover

2. Remove any medicine left in the container

3. Remove the water from the tank

4. Rinse the air nozzle, the extension tube and the mask in warm running water for 30 seconds

5. Clean the main element, the cover and the lower air cover as well as the water tank using a damp 

cloth and a normal detergent

6. Dry using a clean cloth

Once or twice a week

1. Clean the air nozzle, the extension tube and the mask in a much more in-depth manner. Plunge them 

in a solution containing 1 glass of white vinegar and 2 glasses of warm water for 30 minutes.

2. Remove and rinse them.

3. Clean the main unit, the cover, the small air cover and the water tank using a soft cloth with water 

and a normal detergent.

4. Dry using a clean cloth.

OPERATION OF THE AEROSOL

Following the instructions provided above for cleaning before using the ultrasound Aerosol for the first 

time or after a long period in which the device is not used.

Remember: 

Always remove the plug before carrying out cleaning operations and filling with 

every use.

1. The operator should sit in a comfortable and erect position.

2. Connect the aerosol to the plug

3. Press the ignition button at the back of the device in order to activate the stand-by mode.  The green 

light will turn on.

First step: fill the water tank

1. Remove the cover by lifting it, pull out the medicine container if available

2. Use water at a temperature of between 20°C and 30°C (68°F and 86°F), otherwise nebulisation will 

be affected.  The use of distilled water or a physiological solution is preferred as these liquids prevent 

the formation of limestone inside the device that may be ruined if limestone is formed at the base of the 

device, a base that allows for the activation of ultrasounds.

3. Fill the tank with 70 ml of water as described above, until reaching the black plastic base.

NB:  

-The use of cold water (below 20°C or 68°F) would generate a low level of nebulisation

 

-The use of hot water (more than 30°C or 86°F) would generate greater nebulisation

Second step: Fill the medicine container

1. Place the medicine container inside the water tank

2. Pour the medicine directly into the relative container

3. Put the cover back on as well as the air cover and press the stand-by button to start nebulisation 

Remember:

- The maximum capacity of the medicine container is 10 ml and this level should not be 

exceeded

- Only use medicinal products recommended by your doctor