2-14
Installation
2
V
alve Inst
allation or Replacement
Rheodyne 7060 or 9060 valve. To place the valve in
its home position:
1 If the selector knob is installed, remove the knob
by removing the screw that secures the knob to
the valve.
2 Insert a hex wrench into the hole in the valve
shaft and rotate until the flat edge on the shaft
is aligned with port 1.
Rheodyne 7610 valve. To place the valve in its
home position:
1 If the selector knob is installed, remove the knob
by removing the screw that secures the knob to
the valve.
2 Insert a hex wrench into the hole in the valve
shaft and rotate until the flat edge on the shaft
is aligned with port 10.