4-11
Maintenance
4
Replacing a Level Sensing Cable
Removing the level sensing cable
1 Detach the transfer tubing
from the top of the isolation
probe holder by using the
supplied headless nut extender
(part number 49041032) to
loosen and remove the 1/4"-28
nut and ferrule.
2 Unplug the right-angle
connector for the liquid level
sensing cable (LLD cable) from
the socket on the back of the
Z-arm.
3 Detach the LLD cable from the
LLD cable guide.
4 Loosen the small hexagonal
nut on the probe holder for
probe one. Detach the split-
tongue connector from the nut.
5 Remove the transfer tubing
from the LLD cable casing
.
Replacing Parts