Maintenance
4
Cleaning
Biological fluids - Follow this procedure if the
fluid path is in contact with biological fluids such
as blood products:
1 Make a solution of 10% bleach by adding one
part of commercial bleach to nine parts of water.
2 Prime the fluid path using the bleach solution
until the entire fluid path has come into contact
with bleach.
3 Pause the priming sequence.
4 After 30 minutes, resume priming the fluid path
using distilled or deionized water to pump the
remaining bleach solution from the tubing into a
waste container. Prime the fluid path a minimum
of 10 cycles with distilled or deionized water.
5 When you’re satisfied that the entire fluid path
has been flushed with water, end the priming
sequence.