background image

2/6

PIPETMAX ALIGNMENT UTILITY 

|

 INSTRUCTIONS

Run PIPETMAX Alignment Utility

Start Alignment Utility

1.  Click the 

Start

 button and then select: 

(All apps)

 > 

Gilson Applications

 > 

PIPETMAX Alignment Utility

.

The utility starts and then searches for a connected PIPETMAX. If PIPETMAX is not 
connected, a message appears:

Connect PIPETMAX. The utility will automatically scan to search for a connected 
PIPETMAX and then will display the serial number when one is connected.

2.  Select 

Start Alignment. 

A message box appears:

If the alignment head is installed, the alignment cable is connected, the clips are 
positioned correctly, and the rotating cover or fume hood door is closed, select 

Yes

 and 

then proceed t

align pIpetMaX

 on page 5.

If the alignment head is not installed, select 

No

 and then proceed to 

Install alignment head

 on page 3.

If the alignment cable is not connected and/or the clips are not positioned properly, select 

No

 and then go to 

Connect alignment Cable and align Clips

 on page 4.

Reviews: