4/6
PIPETMAX ALIGNMENT UTILITY
|
INSTRUCTIONS
4
Connect Alignment Cable and Align Clips
NOTICE
If the alignment head is installed, but the
cabling is not connected and the alignment
starts, it may damage the alignment head.
1. If your PIPETMAX does not have a rotating cover and
uses the external safety interlock sensor, disconnect
the BOTTOM terminal block connector from the
input/output ports on the rear panel.
2. Pass the end of the alignment cable with pre-wired
connector through the access hole at the back of
PIPETMAX.
3. The connector is labeled
BOTTOM
because it connects
to the bottom set of input/output ports on the rear
panel of PIPETMAX. Make the connection.
4. The clips to secure the tip racks on the tray can
interfere with the alignment. Ensure that the clips (
) are positioned as shown in the
diagram below, and that the thumbscrews have been tightened.
Figure 4
Proper Clip Position During Alignment
5. Close the rotating cover or the fume hood door.
6. Select
OK
to start running the alignment.
Figure 3
Alignment Cable Connection