background image

4/6

PIPETMAX ALIGNMENT UTILITY 

|

 INSTRUCTIONS

4

Connect Alignment Cable and Align Clips

NOTICE

If the alignment head is installed, but the 

cabling is not connected and the alignment 

starts, it may damage the alignment head. 

1.  If your PIPETMAX does not have a rotating cover and 

uses the external safety interlock sensor, disconnect 
the BOTTOM terminal block connector from the 
input/output ports on the rear panel.

2.  Pass the end of the alignment cable with pre-wired 

connector through the access hole at the back of 
PIPETMAX.

3.  The connector is labeled 

BOTTOM

 because it connects 

to the bottom set of input/output ports on the rear 
panel of PIPETMAX. Make the connection.

4.  The clips to secure the tip racks on the tray can 

interfere with the alignment. Ensure that the clips (

) are positioned as shown in the 

diagram below, and that the thumbscrews have been tightened.

Figure 4

Proper Clip Position During Alignment

5.  Close the rotating cover or the fume hood door.
6.  Select 

OK

 to start running the alignment.