5/6
PIPETMAX ALIGNMENT UTILITY
| INSTRUCTIONS
Align PIPETMAX
Wait approximately 6 minutes while the alignment runs.
NOTE
If the rotating cover or fume hood door is opened while running, the alignment is canceled.
A notification appears when the alignment is complete. Remove the alignment head.
Stop Alignment
To stop the alignment while it is running and before it is finished, select
Stop Alignment
. The
alignment will stop when it finishes its current action and revert back to the original offset
values.
Emergency Stop
To stop PIPETMAX immediately, press the red STOP button on the front of PIPETMAX.
Remove Alignment Head
Remove the alignment head by disconnecting the
terminal block connector from the rear panel, and
then removing the thumb nuts securing it to the upper
pipette head assembly. Cover the hole on the rear panel
from the exterior using the black plug supplied in the
accessory package.
black
plug
Figure 6
Black Plug in Access Hole on Rear Panel
Figure 5
Alignment Head Removal