7
PIPETMAN® G
|
QUICK GUIDE
QUICK GUIDE
|
PIPETMAN® G
Ejector spacer
D10
Short tips
Long tips
Eject
or
spac
er
Figure 6
Ejector spacer location
PIPETMAN DIAMOND Tips are
best fitted onto multichannel
pipettes from the patented
ROCKY RACK available only in
our TIPACKS and TOWERPACKS.
Figure 7
ROCKY RACK™ system
Pre-Rinse the Tips
Some liquids (e.g., protein-containing solutions and
organic solvents) can leave a film of liquid on the inside
wall of the tip. Pre-rinse the tip to minimize any errors
that may be related to this phenomenon.
Pre-rinsing consists of aspirating the first volume of
liquid and then dispensing it back into the same vessel
(or to waste). Subsequent volumes you pipette will have
levels of accuracy and precision within specifications.
Ejector Spacer for PIPETMAN G 10 µL
Multichannel Models
According to the tip used, D10 or DL10, you may have to
exchange the ejector spacer. The broadly spaced one is
for D10, and the narrowly spaced one is for DL10.
1.
Remove the tip ejector, keeping both ejector locks
depressed. Pull the tip ejector down.
2. Gently press the tabs from the ejector spacer and
remove it from the tip ejector.
3. Insert the alternative
ejector spacer and click
it to the tip ejector.
4. To refit the tip ejector,
gently re-insert the tip
ejector vertically into the
rails of the ejector
support.