PIPETMAN® G

 

|

 

QUICK GUIDE

QUICK GUIDE 

|

 

PIPETMAN® G

Ejector spacer 

D10

Short tips

Long tips

Eject

or 

spac

er

Figure 6

Ejector spacer location

PIPETMAN DIAMOND Tips are 

best fitted onto multichannel 

pipettes from the patented 

ROCKY RACK available only in 

our TIPACKS and TOWERPACKS.

Figure 7

ROCKY RACK™ system

Pre-Rinse the Tips

Some liquids (e.g., protein-containing solutions and 
organic solvents) can leave a film of liquid on the inside 
wall of the tip. Pre-rinse the tip to minimize any errors 
that may be related to this phenomenon. 

Pre-rinsing consists of aspirating the first volume of 
liquid and then dispensing it back into the same vessel 
(or to waste). Subsequent volumes you pipette will have 
levels of accuracy and precision within specifications.

Ejector Spacer for PIPETMAN G 10 µL 

Multichannel Models

According to the tip used, D10 or DL10, you may have to 

exchange the ejector spacer. The broadly spaced one is 

for D10, and the narrowly spaced one is for DL10.
1. 

Remove the tip ejector, keeping both ejector locks 

depressed. Pull the tip ejector down.

2.  Gently press the tabs from the ejector spacer and 

remove it from the tip ejector.

3.  Insert the alternative 

ejector spacer and click  

it to the tip ejector.

4.  To refit the tip ejector,  

gently re-insert the tip 

ejector vertically into the 

rails of the ejector 

support.

Summary of Contents for PIPETMAN G

Page 1: ...Quick Guide EN G...

Page 2: ...y refer to Fitting the Tips on page 6 Certi cate of Conformity including barcode sticker Multichannel models PIPETMAN G PIPETMAN G Quick Guide Safety bag Ejector spacers for D10 tips for Px10G models...

Page 3: ...and precision The data provided in the Maximum Permissible Errors table on page 5 were obtained using PIPETMAN DIAMOND Tips These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used...

Page 4: ...SYSTEMA TIC ERROR L RANDOM ERROR L PIPETMAN G SINGLE CHANNEL P2G P N F144054M WITH D10 OR DL10 TIPS 0 2 0 5 1 2 0 024 0 025 0 025 0 030 0 012 0 012 0 012 0 014 0 08 0 08 0 08 0 08 0 04 0 04 0 04 0 04...

Page 5: ...e setting slowly reach the required setting making sure to not overshoot the mark When increasing the volume setting pass the required value by 1 3 of a turn and then slowly decrease to reach the volu...

Page 6: ...122 ACCESSORIES Fitting the Tips on P2G and P10G Models NOTE For the P2G and P10G models a dual position adapter plastic is required to fit DL10 tips long tips or D10 tips short tips The metallic rod...

Page 7: ...rst volume of liquid and then dispensing it back into the same vessel or to waste Subsequent volumes you pipette will have levels of accuracy and precision within specifications Ejector Spacer for PIP...

Page 8: ...d touching the tip s orifice First stop Second stop Dispense 1 Place the end of the tip against the inside wall of the recipient vessel at an angle of 10 to 40 2 Press the push button slowly and smoot...

Page 9: ...p ejector tip holder seal and O ring NOTICE PIPETMAN P2G and P10G should not be disassembled You may only replace the push button tip ejector and its dual position adapter With these pipettes if the t...

Page 10: ...or refund of the purchase price paid for such pipettes Routine cleaning control and recalibration are not covered under the warranty The replacement of wearing parts such as seals O rings broken pist...

Page 11: ...11 PIPETMAN G QUICK GUIDE NOTES...

Page 12: ...www gilson com contactus LT801624 C 2019 Gilson Inc All rights reserved...

Reviews: