background image

4

QUICK GUIDE 

|

 

PIPETMAN® G

Maximum Permissible Errors

The data provided in the following table conform to the 

ISO 8655-2 standard. 

 

PIPETMAN G MULTICHANNEL 

P8X10G (P/N F144068) P12X10G (P/N F144069) 

WITH D10 AND DL10 TIPS

1

5

10

± 0.08

± 0.20

± 0.20

 0.05

 0.10

 0.10

± 0.24

± 0.24

± 0.24

 0.16

 0.16

 0.16

P8X20G (P/N F144070) P12X20G (P/N F144071) W

ITH DL10 TIPS

2

10

20

± 0.10

± 0.20

± 0.40

 0.08

 0.10

 0.20

± 0.4

± 0.4

± 0.4

 0.20

 0.20

 0.20

P8X200G (P/N F144072) P12X200G (P/N F144073) 

WITH D200 TIPS

20

100

200

± 0.50

± 1.00

± 2.00

 0.25

 0.40

 0.50

± 3.2

± 3.2

± 3.2

 1.2

 1.2

 1.2

P8X300G (P/N F144074) P12X300G (P/NF144075) 

WITH D300 TIPS

30

150

300

± 1.00

± 1.50

± 3.00

 0.35

 0.60

 1.00

± 8.0

± 8.0

± 8.0

 3.0

 3.0

 3.0

MAXIMUM PERMISSIBLE ERRORS

VOLUME

(µL)

GILSON

ISO 8655

SYSTEMATIC 

ERROR (µL)

RANDOM 

ERROR (µL)

SYSTEMATIC 

ERROR (µL)

RANDOM 

ERROR (µL)

PIPETMAN G SINGLE CHANNEL 

P2G (P/N F144054M) 

WITH D10 OR DL10 TIPS

0.2

0.5

1

2

± 0.024

± 0.025

± 0.025

± 0.030

 0.012

 0.012

 0.012

 0.014

± 0.08

± 0.08

± 0.08

± 0.08

 0.04

 0.04

 0.04

 0.04

P10G (P/N F144055M) 

WITH D10 OR DL10 TIPS

1

5

10

± 0.025

± 0.075

± 0.100

 0.012

 0.030

 0.040

± 0.12

± 0.12

± 0.12

 0.08

 0.08

 0.08

P20G (P/N F144056M) W

ITH D200 TIPS

2

10

20

± 0.10

± 0.10

± 0.20

 0.030

 0.050

 0.060

± 0.2

± 0.2

± 0.2

 0.1

 0.1

 0.1

P100G (P/N F144057M)

  WITH D200 TIPS

10

50

100

± 0.35

± 0.40

± 0.80

 0.10

 0.12

 0.15

± 0.8

± 0.8

± 0.8

 0.3

 0.3

 0.3

P200G (P/N F144058M)

 WITH D200 TIPS

20

100

200

± 0.50

± 0.80

± 1.60

 0.20

 0.25

 0.30

± 1.6

± 1.6

± 1.6

 0.6

 0.6

 0.6

P1000G (P/N F144059M) 

WITH D1000 TIPS

100

500

1000

± 3.0

± 4.0

± 8.0

 0.6

  1.0

  1.5

± 8.0 

± 8.0

± 8.0

 3.0

 3.0

 3.0

P5000G* (P/N F144066) 

WITH D5000 TIPS

500

2500

5000

± 12

± 15

± 30

 3

 5

 8

± 40

± 40

± 40

 15.0

 15.0

 15.0

P10mLG* (P/N F144067) 

WITH D10ML TIPS

1000

5000

10000

± 30

± 40

± 60

 6

 10

 16

± 60

± 60

± 60

 30.0

 30.0

 30.0

* P5000G and P10mLG do not have tip ejectors.

Summary of Contents for PIPETMAN G

Page 1: ...Quick Guide EN G...

Page 2: ...y refer to Fitting the Tips on page 6 Certi cate of Conformity including barcode sticker Multichannel models PIPETMAN G PIPETMAN G Quick Guide Safety bag Ejector spacers for D10 tips for Px10G models...

Page 3: ...and precision The data provided in the Maximum Permissible Errors table on page 5 were obtained using PIPETMAN DIAMOND Tips These values are guaranteed only when genuine PIPETMAN DIAMOND Tips are used...

Page 4: ...SYSTEMA TIC ERROR L RANDOM ERROR L PIPETMAN G SINGLE CHANNEL P2G P N F144054M WITH D10 OR DL10 TIPS 0 2 0 5 1 2 0 024 0 025 0 025 0 030 0 012 0 012 0 012 0 014 0 08 0 08 0 08 0 08 0 04 0 04 0 04 0 04...

Page 5: ...e setting slowly reach the required setting making sure to not overshoot the mark When increasing the volume setting pass the required value by 1 3 of a turn and then slowly decrease to reach the volu...

Page 6: ...122 ACCESSORIES Fitting the Tips on P2G and P10G Models NOTE For the P2G and P10G models a dual position adapter plastic is required to fit DL10 tips long tips or D10 tips short tips The metallic rod...

Page 7: ...rst volume of liquid and then dispensing it back into the same vessel or to waste Subsequent volumes you pipette will have levels of accuracy and precision within specifications Ejector Spacer for PIP...

Page 8: ...d touching the tip s orifice First stop Second stop Dispense 1 Place the end of the tip against the inside wall of the recipient vessel at an angle of 10 to 40 2 Press the push button slowly and smoot...

Page 9: ...p ejector tip holder seal and O ring NOTICE PIPETMAN P2G and P10G should not be disassembled You may only replace the push button tip ejector and its dual position adapter With these pipettes if the t...

Page 10: ...or refund of the purchase price paid for such pipettes Routine cleaning control and recalibration are not covered under the warranty The replacement of wearing parts such as seals O rings broken pist...

Page 11: ...11 PIPETMAN G QUICK GUIDE NOTES...

Page 12: ...www gilson com contactus LT801624 C 2019 Gilson Inc All rights reserved...

Reviews: