CENTRY™ 103 MINICENTRIFUGE
|
USER’S GUIDE
Chapter 4
MAINTENANCE
Routine Maintenance
NOTICE
Ensure that the instrument or any parts of the instrument have been carefully
decontaminated after exposure to infectious, radioactive, or other harmfull substances.
WARNING
Unplug the instrument before performing maintenance procedures. It may cause a
serious electrical shock.
Cleaning and Disinfecting
For cleaning the CENTRY 103 Minicentrifuge, wipe the unit with a soft cloth moistened
with a neutral detergent. The unit can be disinfected with ethanol.
CAUTION
Do not sterilize the rotors by autoclave or with UV, it could damage them. The rotors
can be disinfected with ethanol.
Inspecting and Maintaining
●
Check the rotor, main body, and lid for damage, deformation, and corrosion.
●
Check the lid to make sure it opens and closes smoothly. Press down on the lid to
make sure it is locked into place.
●
Slowly rotate the rotor by hand to ensure it rotates smoothly and does not contact
any other component.
●
Operate the minicentrifuge with no microtubes in the rotor and make sure it is running
safely without excessive vibration.
●
Check the brake for wear. Verify if it takes an abnormally long time for the rotor to
come to a complete stop or if it takes longer to stop than it used to.
●
Check the AC adapter for damage, cracks, or broken cable.
NOTE
Inspection is recommended to be performed on a monthly basis.