CENTRY™ 103 MINICENTRIFUGE

 

|

 

USER’S GUIDE

Chapter 2

SAFETy AND USER pRECAUTIONS 

Read the

 User’s Guide carefully before operating the instrument and follow all safety norms.

General Safety Precautions

NOTICE

When the CENTRy 103 Minicentrifuge is not in use, ensure the power is switched off  

(position “O”). When the power is switched on (position “I”), the red, indicator light turns on.

When disposing of the instrument, please follow the relevant laws and local or state 

regulations for the waste disposal. 

Rotor and Microtubes

NOTE

After the operation, remove the microtubes immediately from the rotor. If the tubes 

are left inside the chamber after a run, the remaining heat of the motor may increase 

the temperature of the samples.

WARNING

Do not modify or alter the instrument or any parts of the instrument, it may cause 

serious accident or malfunction.
Do not use accessories or parts other than those specified in this 

User’s Guide, which 

may cause serious accident or malfunction.
Do not touch the power plug with a wet hand, it may cause a serious or deadly 

electrical shock.
Do not use any damaged power cord, plug, or loose outlet. It may cause an 

accidental fire, short circuit, or malfunction.
Never open the lid while the rotor is spinning, accidental contact with the spinning 

rotor may cause serious injury.

When leaving the instrument out of use for a long period of time, be sure to 

remove the power plug from the wall outlet for safety. Otherwise the insulation 

could deteriorate, thereby causing an electrical shock, short circuit, or fire.

WARNING

Ensure the rotor is securely installed, otherwise it may fall off during spinning and 

cause unexpected accidents, or injury.
Do not use any damaged, deformed, or corroded centrifuge microtubes. Ensure 

that the microtubes used are compatible with the maximum RCF of the instrument, 

otherwise they could break while spinning and cause a serious injury.
The load of the microtubes must be appropriately balanced. By applying excessive 

force to the motor shaft, unbalanced load may cause damage to the instrument, 

resulting in a serious injury.

 Batteries

NOTE

As the battery life decreases, the rotating speed also decreases. For an optimal use of 

the instrument, replace the batteries with new ones.

WARNING

For using the instrument with batteries, ensure that the positive (+) and the negative 

(-) terminals are correctly oriented, as represented in the battery compartment. 

Otherwise, the battery may cause heating, leakage, or explosion and could result in 

damage to the instrument, or personal injury.

Summary of Contents for CENTRY 103 Minicentrifuge

Page 1: ...User s Guide EN CENTRY 103 Minicentrifuge...

Page 2: ...and Microtubes 5 Batteries 5 OPERATION 6 Installing 6 Connecting to the Power Source and Turning On 6 Installing the Rotor 7 Preparing the Microtubes 8 Starting and Ending Operation 8 Using Batteries...

Page 3: ...in the lid The CENTRY 103 Minicentrifuge will automatically decelerate to a stop after six minutes of continuous spinning Ordering Information DESCRIPTION PART NUMBER CENTRY 103 Minicentrifuge F11073...

Page 4: ...AC adapter connector 7 Battery compartment 4 AA size LR6 alkaline batteries 1 5V 8 Rotor for microtubes 8 microtubes of 1 5 mL or 2 0 mL 9 Rotor for tube strips 2 lines of 8 connected 0 2 mL microtub...

Page 5: ...ntal fire short circuit or malfunction Never open the lid while the rotor is spinning accidental contact with the spinning rotor may cause serious injury When leaving the instrument out of use for a l...

Page 6: ...unlight Avoid installing the minicentrifuge on a slippery surface or surface prone to vibration WARNING Do not install the instrument in facilities that store chemicals that could generate flammable g...

Page 7: ...locked inside the hole of the rotor For dismounting the rotor hold it by both hands and push the tabs inward with two thumbs to disconnect them from the rotor and then pull the rotor out from the moto...

Page 8: ...ng run Starting and Ending Operation By closing the lid 1 the rotor instantly starts spinning After six minutes of spinning the rotor will automatically decelerate to a stop WARNING Ensure that there...

Page 9: ...oclave or with UV it could damage them The rotors can be disinfected with ethanol Inspecting and Maintaining Check the rotor main body and lid for damage deformation and corrosion Check the lid to mak...

Page 10: ...t is activated Turn off the power switch and wait at least one minute The rotor stops while it is spinning Motor s overcurrent is detected by unbalanced load and the overcurrent protection circuit is...

Page 11: ...R S GUIDE Chapter 5 SPARE PARTS Accessories DESCRIPTION PART NUMBER CENTRY 103 ROTOR 8 TUBES 1 5 2 0 mL F1077310 CENTRY 103 ROTOR 2 TUBE STRIPS 8 X 0 2 mL F1077311 CENTRY 103 MOTOR SHAFT BOSS F1077312...

Page 12: ...voltage equipment 2014 35 EU Electromagnetic compatibility 2014 30 EU Restriction of hazardous substances 2011 65 EU Standards for EU Safety requirements for electrical equipment for measurement cont...

Page 13: ...only if the instrument is used in the manner described in this guide and for the purpose for which it is designed Gilson is not responsible for consequential damages resulting from the misuse or bad c...

Page 14: ...www gilson com contactus LT801606 A 03 2019 Gilson Inc All rights reserved...

Reviews: