CENTRY™ 103 Minicentrifuge is an easy of use and portable benchtop device ideal for a 

quick separation of particles from a supernatant according to a density gradient.  

It also enables a spin down of a drop of liquid from the walls of tubes.
Equipped with robust and easy to change rotors, the CENTRY 103 Minicentrifuge can be 

used either with microtubes or tube-strips. The round rotor has a capacity of eight  

1.5 mL or 2.0 mL microtubes, and the rectangular shaped rotor capacity is two lines of 

eight connected 0.2 mL tube-strips. 
Due to its small footprint and a battery compartment, the CENTRY 103 Minicentrifuge 

can be used anywhere, from small benches to laminar flow cabinets, or even if there is no 

power supply available, the Minicentrifuge is convenient and easy to transport. 
To start or stop centrifugation, simply press down in the lid.  

The CENTRY 103 Minicentrifuge will automatically decelerate to a stop after six minutes of 

continuous spinning.

Ordering Information

DESCRIPTION 

PART NUMBER 

CENTRY™ 103 Minicentrifuge

F110736

Specifications 

ROTOR 8 TUBES  

1.5–2.0 mL (PN F1077310)

ROTOR 2 TUBE-STRIPS  

8 x 0.2 mL (PN F1077311)

Dimensions  

(W X D X H)

150 x 163 x 113 mm

Weight

600 g (excluding rotor and batteries)

Max. Capacity

8 microtubes: 
2.0 mL or 1.5 mL

2 PCR strips:
0.2 mL x 8 connected tubes

Speed

6 000 rpm (+/- 20%)

6 200 rpm (+/- 20%)

Relative Centrifugal 
Force (RCF)

1.310 ~ 2.960G (2.0 mL)
1.260 ~ 2.840G (1.5 mL)

Max. 1.200 ~ 2.690G
Min. 810 ~ 1.820G

Safety

Automatic stop after six minutes (fixed timer)
Lid opening interlocks brake
Overcurrent fuse and motor overcurrent detector  
(Motor locks when the rotor is unbalanced)

Power

AC adapter with five different plugs (1 SKU)
Input: 1-Phase AC 100-240V | 50/60Hz | 0.6A
Or four AA alkaline batteries

Autonomy with 

batteries (approx. 

indication)

 

about 170 cycles of  

10 seconds spinning

 

about 7 cycles of  

6 minutes spinning

 

about 500 cycles of  

10 seconds spinning

 

about 15 cycles of 6 minutes 

spinning

Warranty 

12 months

Summary of Contents for CENTRY 103 Minicentrifuge

Page 1: ...User s Guide EN CENTRY 103 Minicentrifuge...

Page 2: ...and Microtubes 5 Batteries 5 OPERATION 6 Installing 6 Connecting to the Power Source and Turning On 6 Installing the Rotor 7 Preparing the Microtubes 8 Starting and Ending Operation 8 Using Batteries...

Page 3: ...in the lid The CENTRY 103 Minicentrifuge will automatically decelerate to a stop after six minutes of continuous spinning Ordering Information DESCRIPTION PART NUMBER CENTRY 103 Minicentrifuge F11073...

Page 4: ...AC adapter connector 7 Battery compartment 4 AA size LR6 alkaline batteries 1 5V 8 Rotor for microtubes 8 microtubes of 1 5 mL or 2 0 mL 9 Rotor for tube strips 2 lines of 8 connected 0 2 mL microtub...

Page 5: ...ntal fire short circuit or malfunction Never open the lid while the rotor is spinning accidental contact with the spinning rotor may cause serious injury When leaving the instrument out of use for a l...

Page 6: ...unlight Avoid installing the minicentrifuge on a slippery surface or surface prone to vibration WARNING Do not install the instrument in facilities that store chemicals that could generate flammable g...

Page 7: ...locked inside the hole of the rotor For dismounting the rotor hold it by both hands and push the tabs inward with two thumbs to disconnect them from the rotor and then pull the rotor out from the moto...

Page 8: ...ng run Starting and Ending Operation By closing the lid 1 the rotor instantly starts spinning After six minutes of spinning the rotor will automatically decelerate to a stop WARNING Ensure that there...

Page 9: ...oclave or with UV it could damage them The rotors can be disinfected with ethanol Inspecting and Maintaining Check the rotor main body and lid for damage deformation and corrosion Check the lid to mak...

Page 10: ...t is activated Turn off the power switch and wait at least one minute The rotor stops while it is spinning Motor s overcurrent is detected by unbalanced load and the overcurrent protection circuit is...

Page 11: ...R S GUIDE Chapter 5 SPARE PARTS Accessories DESCRIPTION PART NUMBER CENTRY 103 ROTOR 8 TUBES 1 5 2 0 mL F1077310 CENTRY 103 ROTOR 2 TUBE STRIPS 8 X 0 2 mL F1077311 CENTRY 103 MOTOR SHAFT BOSS F1077312...

Page 12: ...voltage equipment 2014 35 EU Electromagnetic compatibility 2014 30 EU Restriction of hazardous substances 2011 65 EU Standards for EU Safety requirements for electrical equipment for measurement cont...

Page 13: ...only if the instrument is used in the manner described in this guide and for the purpose for which it is designed Gilson is not responsible for consequential damages resulting from the misuse or bad c...

Page 14: ...www gilson com contactus LT801606 A 03 2019 Gilson Inc All rights reserved...

Reviews: