32 Auto Ref/Keratometer GRK-7000
-------------------------------------------------------------
n by slowly pushing and rotating the operation lever forward. At this tim
e, let the glittering bright dot come into the core of internal array ring.
-
Ask the examinee to look at the internal fixed target.
-
Adjust the focus so that the outline of bright dot can be apparent. If the
focus is adjusted appropriately, the circle symbol appears on the bright
dot.
-
Height Adjustment: Adjust it by rotating t
he operation lever or chin-rest lever.
-
Left/Right Adjustment: Move the operatio
n lever left and right so that the Outer A
lignment Ring is aligned with the Mire Im
age
-
Focus Adjustment: Adjust it to the bright dot by tilting the operation leve
r forward/backward.
[ Figure 14. Height Adjustment
]