Auto Ref/Keratometer GRK-7000 43
8.2. Keratometry (KER Mode)
It is the mode to measure the corneal curvature solely.
Do not measure the base curve of hard contact lens in this mode. Please refer
to CLBC mode in section 8.4 regarding the base curve of hard contact lens.
1. Check whether or not the screen of monitor is in measurement mode.
2. KER Mode Selection
-Push MODE button until “KER” is to be indicated on the upper right side
of the screen.
3. Perform the same 2, 3 procedure of consecutive measurement of refractive p
ower.
8.2.1.
Manual Measurement Mode
① Perform the adjustment of array and focus as like in the procedure ○
1
, ○
2
of
section 8.1.1.
② -Measurement
-
Push the measurement button.
-
The measurement continues to be performed as you keep pushing the m
easurement button.
-
As the measurement is completed, the measured result is to be indicated
Summary of Contents for GRK-7000
Page 1: ...User s Manual Auto Ref Keratometer GRK 7000...
Page 39: ...38 Auto Ref Keratometer GRK 7000 Figure 17 Example of Print...
Page 42: ...Auto Ref Keratometer GRK 7000 41...
Page 47: ...46 Auto Ref Keratometer GRK 7000 Figure 20 Example of Print...
Page 54: ...Auto Ref Keratometer GRK 7000 53...
Page 57: ...56 Auto Ref Keratometer GRK 7000 7 Print Press Print button...