Operation
MetPak II User Manual
Last Change
©
Gill Instruments Ltd
20 August, 2010
Page 23
3.2
Configuring MetPak II
MetPak II can be configured using:
A terminal program such as MS Windows HyperTerminal.
Gill Wind software used as a Terminal Program only.
The default configuration settings are listed in the following table.
Table 8
Default configuration settings
Setting
Function
Comms Levels
RS232
Baud Rate
19200
Data Connection
Data Bits 8, Parity None, Stop Bits 1, Flow Control None.
Output rate
1Hz
Message Mode
Continuous
Message Format
Wind Direction (degrees), Wind Speed (metres/Sec),
Barometric Pressure (hecto pascals), Humidity (percent),
Temperature (degrees C), Dew Point (degrees C), Power
Supply (volts), Status Code (00) Checksum (hex number).
Output Sentence
Comma Separated Variable
Node Address
Q
Message terminator
Carriage return and line feed.
Power up message
On
Example Default Data String
:
<STX>Q,170,000.08,1019.5,035.0,+024.7,+008.3,+04.9,00,<ETX>55 & (CR,LF)
STX
Node letter
Wind direction
Wind speed
Pressure
Humidity
Temperature
Dewpoint
Supply voltage
Status code
ETX
Checksum
Carriage Return/Line Feed
NOTES:
<STX> is the Start of String character (ASCII value 2).
<ETX> is the End of String character (ASCII value 3).
Checksum, the 2 digit Hex Checksum sum figure is calculated from the Exclusive OR of the bytes
between (and not including) the STX and ETX characters.