Page
9
2 Device description P
-
9710 optometer
2.3 Technical specifications P
-
9710 series
Power supply
built
-
in lead acid battery, 6V / 0.5Ah (Version before
2017)
Built
-
in lithium Ion battery, 3.7V /1400mAh (Version
2017)
operating time: 5hr / 12hr with backlight on /off
plug 5.5mm/2.5mm/9.5mm for external power
supply (7.5VDC
-
8.0VDC / min. 250mA)
Display
alphanumeric LCD, 2 rows x 16 chars
character height 5mm
LED backlight (on / off)
Detector interface
9pin DSUB (female)
Analog output
output voltage proportional to detector input current
connector: TRIAD01 5pin (Tyco)
Units of measurement
Amperes (absolute calibration)
optical units corresponding to calibration data
ratio in percent, logarithmic, and as factor
Signal amplifier
Transimpedance amplifier
8 ranges (200pA to 2mA)
fixed range or autorange
CW integration time
100µs
-
5.9999s
Pulse integration time
10ms
-
199.99s
Offset correction
correction range transcending
Parameter adjustment
front panel keys or remote control
adjusted values permanently stored in EEPROM
10 different overall settings can be stored
Optical calibration data
stored in detector head connector (EEPROM)
max. 250 calibration table entries
interpolation between wavelength points possible
manual input of a calibration factor possible
Data logger
max. 12288 entries, permanently stored in flash
memory
Front panel control
10 buttons, menu system
Remote interface
RS232 (9600 Baud, 8 Data Bits, 1 Stop Bit, No
Parity)
Connector: TRIAD01 (Tyco)