Page
37
A Appendix
-
Measurement Modes
set menu "Setup" / "Trigger Input" to "active"
set menu “Setup” / “Synchronisation” to “Not active”
set menu “Setup” / “Integration Time” to “0.3000s”
set menu “Setup” / “Zero Measurement” to “Fixed Values”
set menu “Setup” / “Pulse Meas.
-
Time” to “0.200s”
set menu “Setup” / “Pulse Offset” to “Continuous”
Initiate Measurement:
press <run / stop> key to activate measurement
set trigger input to low level to start measurement
trigger pulse source (immediately after or simultaneously with setting trig-
ger input to low level)
read measurement result on P
-
9710 display
if “UDL” or “OVL” is displayed repeat measurement until a valid measure-
ment result is displayed
Example
-
Pulse Energy Measurement of an Infinite Pulse Chain (pulse
length: 10µs, pulse frequency: 50Hz)
Set P
-
9710 Parameters:
set menu “Mode” to “Pulse Energy”
set menu “Range” to “Auto”
set menu “Setup” / “Synchronisation” to “Not active”
set menu “Setup” / “Integration Time” to “0.3000s”
set menu “Setup” / “Zero Measurement” to “Fixed Values”
set menu “Setup” / “Pulse Meas.
-
Time” to “5.000s”
set menu “Setup” / “Pulse Offset” to “Static”
remove pulse signal from detector
select menu “Offset”, set “Offset = CW”
adjust pulse signal to detector
Initiate Measurement:
press <run / stop> key to start measurement
read measurement result on P
-
9710 display
if “UDL” or “OVL” is displayed repeat measurement until a valid measure-
ment result is displayed