![GHM GREISINGER GIA 20 EB Mounting And Operating Manual Download Page 18](http://html.mh-extra.com/html/ghm/greisinger-gia-20-eb/greisinger-gia-20-eb_mounting-and-operating-manual_2220443018.webp)
E31.0.12.6C-06
Mounting and operating manual GIA 20 EB with electrically insulated supply
page 18 of 32
6.4. Measuring of frequency (TTL, switching-contact)
This chapter describes how to configure the device for measuring frequency.
This instruction demands that you selected “
FrEq
“ as your desired input type like it is explained in
chapter 6.1.
The display must show
INP
.
Parameter
Button 1
Value
Button 2 / 3
Notes
SENS
TTL
TTL-signal (see info connection)
NPN
Switching contact, NPN
For direct connection of a passive switching contact (e.g. push
button, relay) or transmitter with NPN output.
There is a internal pull-up-resistor.
(see info switching contact)
PNP
Switching contact, PNP
For direct connection of a transmitter with PNP output.
There is a internal pull-down-resistor.
FR.LO
0 ... 9999
lower frequency range
FR.K,
FR.LO ... 9999
upper frequency range
DP
"- - - -"
"- - - . -"
"- - . - -"
"- . - - -"
Decimal point
D,.LO
-1999 ... 9999
display at lower frequency range limit
D,.K,
-1999 ... 9999
display at upper frequency range limit
L ,
(Limit)
OFF
Exceeding of the measuring-frequency is tolerable until you
reach the maximum measuring range limit.
ON.ER
The measuring range is exactly bounded by the selected fre-
quency-measuring-range-limit. When exceeding or short falling
of the limit the device will display an error message.
ON.RG
The measuring range is exactly bounded by the selected meas-
uring-range-limit. When exceeding or short falling of the limit the
device will display the lower or upper display-range-limit.
e.g. humidity:
when exceeding, the device will display 0% or 100%
F,LT
0.01 ... 2.00
s
Filter
O
V
TP
Output configuration will be explained in chapter 7. According to
your settings different parameters will be shown.
Finish configuration:
Press
Button 1
as often till the device executes a reset. The values will be stored at it.
.
Switching contact, NPN:
when using push-buttons or relays, they must be bounce-free!
Connection:
For the connection of a frequency-transmitter, please follow the instructions given in chapter 5.3.6.
When connecting a switching-contact-transmitter with increased frequency range (= with external
circuitry) you have to select TTL as input signal.