LPUVA02 

-   3   -

 

V2.0

 

1

 

 INTRODUCTION 

The 

LPUVA02 

radiometer  measures  the  global  irradiance  in  the  UVA  spectral  range 

on a flat surface (W/m

2

). Measured irradiance is the result of the sum of direct solar 

irradiance and of diffuse irradiance.  

The radiometer can also be used for monitoring indoor UVA emissions. 

Available in the following versions: 

 

LPUVA02

PASSIVE 

 

LPUVA02AC

ACTIVE with 4..20 mA CURRENT output (0…200 W/m

2

 

LPUVA02AV

ACTIVE with 0..1 or 0..5 or 0..10 V VOLTAGE output to be de-
fined when ordering (0…200 W/m

2

). 

Each radiometer is individually calibrated at the factory and is distinguished by its own 
calibration factor. The calibration is performed by comparison with the primary stand-
ard  of  the  Delta  OHM  metrology  laboratory,  using  the  emission  line  at  365  nm  of  a 
suitably filtered Xe-Hg lamp. 

Note: there is no international standard for the calibration of UVA radiometers; there-
fore, when considering the value of the calibration coefficient, the method by which it 
was obtained must be taken into account; the same radiometer calibrated with differ-
ent procedures can have different calibration factors. 

Summary of Contents for DeltaOHM LPUVA02

Page 1: ...Operating manual UVA Radiometer LPUVA02 www deltaohm com English Keep for future reference...

Page 2: ...ON 5 4 ELECTRICAL CONNECTIONS 8 4 1 LPUVA02 CONNECTIONS 8 4 2 LPUVA02AC CONNECTIONS 9 4 3 LPUVA02AV CONNECTIONS 9 5 MEASUREMENT 10 5 1 LPUVA02 10 5 2 LPUVA02AC 10 5 3 LPUVA02AV 10 6 MAINTENANCE 11 7 T...

Page 3: ...VOLTAGE output to be de fined when ordering 0 200 W m2 Each radiometer is individually calibrated at the factory and is distinguished by its own calibration factor The calibration is performed by com...

Page 4: ...ined thanks to the par ticular shape of the diffuser and of the housing The deviation between the theoretical response and the measured one is shown in the Fig 2 2 Angle Fig 2 2 The excellent relation...

Page 5: ...a coin 3 Remove the cartridge perforated cap 4 Open the sachet containing silica gel supplied with the radiometer 5 Fill the cartridge with the silica gel crystals 6 Close the cartridge with its own c...

Page 6: ...lf For fixing use the holes on the radiometer body remove the shade disk to access the holes and reposition it after mounting or the suitable accessories see the fig ures below In order to allow an ac...

Page 7: ...LPUVA02 7 V2 0 HD2003 79K HD2003 77C HD2003 83 HD2003 83 1 HD2003 85K Fig 3 3 fixing accessories...

Page 8: ...to the connector 4 1 LPUVA02 CONNECTIONS The radiometer LPUVA02 is passive and does not require power supply It is to be connected either to a millivoltmeter or to a data acquisition system Typically...

Page 9: ...rdered output and requires external power supply 10 30 Vdc for 0 1 V and 0 5 V outputs 15 30 Vdc for 0 10 V output It is to be connected to a power sup ply and an instrument with voltage input as show...

Page 10: ...PUVA02AC The 4 20 mA output signal corresponds to 0 200 W m2 irradiance range The irradiance Ee is obtained by measuring with a multimeter the current Iout absorbed by the sensor and applying the foll...

Page 11: ...is case the performed reading is highly over estimated To minimize the condensation the radiometer is provided with a car tridge containing dessicant material silica gel The efficiency of the silica g...

Page 12: ...93 nm 1 10 320 nm 400 nm 1 100 Peak 365 nm Response time 0 5 s 95 Output LPUVA02 V Wm 2 LPUVA02AC 4 20 mA LPUVA02AV 0 1 0 5 0 10 V depending on the model Power supply 10 30 Vdc 15 30 Vdc only for 0 10...

Page 13: ...safety can be ensured only if all standard safety measures as well as the specific measures described in this manual are followed Do not use the instruments in places where there are Corrosive or flam...

Page 14: ...the radiometer on 40 mm clamping To install the radiometer on a transverse mast LPS6 Kit for the installation of the radiometer The kit includes 750 mm mast base fitting graduated support plate brack...

Page 15: ......

Page 16: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Reviews: