GHM DeltaOHM LPUVA02 Operating Manual Download Page 8

 

LPUVA02 

-   8   -

 

V2.0

 

4

 

ELECTRICAL CONNECTIONS 

LPUVA02… 

has a 4-pole connector and uses the 

CPM12AA4...

 

optional

 cables. 

The  metallic  housing  of  the  radiometer  should  preferably  be 
grounded ( ) locally. In this case, do not connect the wire of the 
cable corresponding to the housing to prevent ground loops. 

Only if it is not possible to ground locally the metallic case of the 
radiometer,  connect  the  wire  of  the  cable  corresponding  to  the 
housing  to  ground.

 Note: in LPUVA02AV the housing is not connected 

to the connector. 

4.1

 

LPUVA02

 CONNECTIONS

 

The  radiometer  LPUVA02  is

  passive

  and  does  not  require  power  supply.  It  is  to  be 

connected either to a millivoltmeter or to a data acquisition system. Typically, the ra-
diometer output signal does not exceed 50 mV. In order to better exploit the radiome-
ter features, the readout instrument should have 1 

µ

V resolution. 

Connector 

Function 

Color 

+Vout 

Red 

-Vout 

Blue 

Housing (C) 

White 

Cable shield (SH) 

Black 

1

2

3

4

+Vout

-Vout

SH

C

 

Fig. 4.1: LPUVA02 connections 

CPM12AA4… cable 

Radiometer M12 

male connector 

Datalogger or 

converter/amplifier 

with 

V

 or 

mA

 output 

Connect to ground only if it is not possible to 
ground locally the housing of the radiometer 

Summary of Contents for DeltaOHM LPUVA02

Page 1: ...Operating manual UVA Radiometer LPUVA02 www deltaohm com English Keep for future reference...

Page 2: ...ON 5 4 ELECTRICAL CONNECTIONS 8 4 1 LPUVA02 CONNECTIONS 8 4 2 LPUVA02AC CONNECTIONS 9 4 3 LPUVA02AV CONNECTIONS 9 5 MEASUREMENT 10 5 1 LPUVA02 10 5 2 LPUVA02AC 10 5 3 LPUVA02AV 10 6 MAINTENANCE 11 7 T...

Page 3: ...VOLTAGE output to be de fined when ordering 0 200 W m2 Each radiometer is individually calibrated at the factory and is distinguished by its own calibration factor The calibration is performed by com...

Page 4: ...ined thanks to the par ticular shape of the diffuser and of the housing The deviation between the theoretical response and the measured one is shown in the Fig 2 2 Angle Fig 2 2 The excellent relation...

Page 5: ...a coin 3 Remove the cartridge perforated cap 4 Open the sachet containing silica gel supplied with the radiometer 5 Fill the cartridge with the silica gel crystals 6 Close the cartridge with its own c...

Page 6: ...lf For fixing use the holes on the radiometer body remove the shade disk to access the holes and reposition it after mounting or the suitable accessories see the fig ures below In order to allow an ac...

Page 7: ...LPUVA02 7 V2 0 HD2003 79K HD2003 77C HD2003 83 HD2003 83 1 HD2003 85K Fig 3 3 fixing accessories...

Page 8: ...to the connector 4 1 LPUVA02 CONNECTIONS The radiometer LPUVA02 is passive and does not require power supply It is to be connected either to a millivoltmeter or to a data acquisition system Typically...

Page 9: ...rdered output and requires external power supply 10 30 Vdc for 0 1 V and 0 5 V outputs 15 30 Vdc for 0 10 V output It is to be connected to a power sup ply and an instrument with voltage input as show...

Page 10: ...PUVA02AC The 4 20 mA output signal corresponds to 0 200 W m2 irradiance range The irradiance Ee is obtained by measuring with a multimeter the current Iout absorbed by the sensor and applying the foll...

Page 11: ...is case the performed reading is highly over estimated To minimize the condensation the radiometer is provided with a car tridge containing dessicant material silica gel The efficiency of the silica g...

Page 12: ...93 nm 1 10 320 nm 400 nm 1 100 Peak 365 nm Response time 0 5 s 95 Output LPUVA02 V Wm 2 LPUVA02AC 4 20 mA LPUVA02AV 0 1 0 5 0 10 V depending on the model Power supply 10 30 Vdc 15 30 Vdc only for 0 10...

Page 13: ...safety can be ensured only if all standard safety measures as well as the specific measures described in this manual are followed Do not use the instruments in places where there are Corrosive or flam...

Page 14: ...the radiometer on 40 mm clamping To install the radiometer on a transverse mast LPS6 Kit for the installation of the radiometer The kit includes 750 mm mast base fitting graduated support plate brack...

Page 15: ......

Page 16: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Reviews: