MAINTENANCE 

In order to grant measurements high accuracy, it is important to keep the glass dome 
clean. Consequently, the more the dome will be kept clean, the more measurements 
will be accurate.  

You  can  wash  it  using  water  and  standard  papers  for  lens.  If  necessary,  use  pure 
ETHYL alcohol. After using alcohol, clean again the dome with water only.  

Because of the high temperature changes between day and night, some condensation 
might  appear  on  the  radiometer  dome.  In  this  case  the  performed  reading  is  highly 
over-estimated. To minimize the condensation, the radiometer is provided with a car-
tridge containing dessicant material (silica-gel). The efficiency of the silica-gel crystals 
decreases  over  time  while  absorbing  humidity.  Silica-gel  crystals  are  efficient  when 
their colour is 

yellow

, while they turn 

white/translucent

 as soon as they lose their 

efficiency. Read instructions at chapter 3 about how to replace the silica-gel crystals. 
Silica-gel  typical  lifetime  goes  from  2  to  6  months  depending  on  the  environment 
where the radiometer works.  

To exploit all the radiometer features, it is highly recommended that the calibration be 
checked annually. 

 

Summary of Contents for DeltaOHM LPUVA02

Page 1: ...Operating manual UVA Radiometer LPUVA02 www deltaohm com English Keep for future reference...

Page 2: ...ON 5 4 ELECTRICAL CONNECTIONS 8 4 1 LPUVA02 CONNECTIONS 8 4 2 LPUVA02AC CONNECTIONS 9 4 3 LPUVA02AV CONNECTIONS 9 5 MEASUREMENT 10 5 1 LPUVA02 10 5 2 LPUVA02AC 10 5 3 LPUVA02AV 10 6 MAINTENANCE 11 7 T...

Page 3: ...VOLTAGE output to be de fined when ordering 0 200 W m2 Each radiometer is individually calibrated at the factory and is distinguished by its own calibration factor The calibration is performed by com...

Page 4: ...ined thanks to the par ticular shape of the diffuser and of the housing The deviation between the theoretical response and the measured one is shown in the Fig 2 2 Angle Fig 2 2 The excellent relation...

Page 5: ...a coin 3 Remove the cartridge perforated cap 4 Open the sachet containing silica gel supplied with the radiometer 5 Fill the cartridge with the silica gel crystals 6 Close the cartridge with its own c...

Page 6: ...lf For fixing use the holes on the radiometer body remove the shade disk to access the holes and reposition it after mounting or the suitable accessories see the fig ures below In order to allow an ac...

Page 7: ...LPUVA02 7 V2 0 HD2003 79K HD2003 77C HD2003 83 HD2003 83 1 HD2003 85K Fig 3 3 fixing accessories...

Page 8: ...to the connector 4 1 LPUVA02 CONNECTIONS The radiometer LPUVA02 is passive and does not require power supply It is to be connected either to a millivoltmeter or to a data acquisition system Typically...

Page 9: ...rdered output and requires external power supply 10 30 Vdc for 0 1 V and 0 5 V outputs 15 30 Vdc for 0 10 V output It is to be connected to a power sup ply and an instrument with voltage input as show...

Page 10: ...PUVA02AC The 4 20 mA output signal corresponds to 0 200 W m2 irradiance range The irradiance Ee is obtained by measuring with a multimeter the current Iout absorbed by the sensor and applying the foll...

Page 11: ...is case the performed reading is highly over estimated To minimize the condensation the radiometer is provided with a car tridge containing dessicant material silica gel The efficiency of the silica g...

Page 12: ...93 nm 1 10 320 nm 400 nm 1 100 Peak 365 nm Response time 0 5 s 95 Output LPUVA02 V Wm 2 LPUVA02AC 4 20 mA LPUVA02AV 0 1 0 5 0 10 V depending on the model Power supply 10 30 Vdc 15 30 Vdc only for 0 10...

Page 13: ...safety can be ensured only if all standard safety measures as well as the specific measures described in this manual are followed Do not use the instruments in places where there are Corrosive or flam...

Page 14: ...the radiometer on 40 mm clamping To install the radiometer on a transverse mast LPS6 Kit for the installation of the radiometer The kit includes 750 mm mast base fitting graduated support plate brack...

Page 15: ......

Page 16: ...rranty included in the manual of the product For any dispute the competent court is the Court of Padua The Italian law and the Convention on Contracts for the International Sales of Goods apply TECHNI...

Reviews: