![GHM DeltaOHM HD2301.0 Operating Manual Download Page 22](http://html1.mh-extra.com/html/ghm/deltaohm-hd2301-0/deltaohm-hd2301-0_operating-manual_2220415022.webp)
HD2301
- 22 -
V1.7
T
ECH N I CAL D ATA OF PROBES AN D M OD ULES I N LI N E W I TH TH E I N STRUM EN T
T
EM PERATURE PROBES
Pt 1 0 0
SEN SOR USI N G
SI CRAM
M OD ULE
M ode l
Ty pe
Applica t ion
r a n ge
Accu r a cy
TP4 7 2 I
I m m ersion
196 ° C…+ 500 ° C
± 0.1° C ( @ 0 ° C)
± 0.2° C ( -
50 °C ≤ t ≤ 250 °C)
± 0.3° C (t < -50 ° C; t > 250 ° C)
TP4 7 2 I .O
I m m ersion
- 50 ° C…+ 300 ° C
TP4 7 3 P.I
Penet rat ion
- 50 ° C…+ 400 ° C
TP4 7 3 P.O
Penet rat ion
- 50 ° C…+ 300 ° C
TP4 7 4 C.O
Cont act
- 50 ° C…+ 300 ° C
TP4 7 5 A.O
Air
- 50 ° C…+ 250 ° C
TP4 7 2 I .5
I m m ersion
- 50 ° C…+ 400 ° C
TP4 7 2 I .1 0
I m m ersion
- 50 ° C…+ 400 ° C
TP4 9 A.I
I m m ersion
- 70 ° C…+ 250 ° C
TP4 9 AC.I
Cont act
- 70 ° C…+ 250 ° C
TP4 9 AP.I
Penet rat ion
- 70 ° C…+ 250 ° C
TP8 7 5 .I
Globe- t herm om et er
Ø 150 m m
- 30 ° C…+ 120 ° C
TP8 7 6 .I
Globe- t herm om et er
Ø 50 m m
- 30 ° C…+ 120 ° C
TP8 7 .O
I m m ersion
- 50 ° C…+ 200 ° C
Com m on charact erist ics
Re solut ion
0 .0 1 ° C in t h e r a n ge ± 1 9 9 .9 9 ° C,
0 .1 ° C in t h e r e m a inin g fie ld
Tem perat ure drift @ 20° C
0.003% / ° C
Pt 1 0 0 / Pt 1 0 0 0
PROBES W I TH
TP4 7
CON N ECTOR W I TH OUT
SI CRAM
M OD ULE
M ode l
Ty pe
Applica t ion r a n ge
Accu r a cy
TP4 7 .1 0 0 .O
Pt 100 4 w ires
- 50…+ 250 ° C
Class A
TP4 7 .1 0 0 0 .O
Pt 1000 4 w ires
- 50…+ 250 ° C
Class A
TP8 7 .1 0 0 .O
Pt 100 4 w ires
- 50…+ 200 ° C
Class A
TP8 7 .1 0 0 0 .O
Pt 1000 4 w ires
- 50…+ 200 ° C
Class A
Com m on charact erist ics
Re solut ion
0 .0 1 ° C in t h e r a n ge ± 1 9 9 .9 9 ° C,
0 .1 ° C in t h e r e m a inin g fie ld
Tem perat ure drift @ 20° C
Pt 100 0.003% / ° C