![GHM DeltaOHM HD2301.0 Operating Manual Download Page 15](http://html1.mh-extra.com/html/ghm/deltaohm-hd2301-0/deltaohm-hd2301-0_operating-manual_2220415015.webp)
HD2301
-
15
-
V1.7
8. I nsert t he gasket in t he m odule;
9. Screw t he fair lead and t he r ing.
10.At t he ot her end of t he m odule, ent er t he r ing w it h t he O- Ring as indicat ed in t he
pict ure.
11.Make sure t he cable is not t w ist ed while you are screwing t he fair lead. Now t he
probe is ready.
D
I RECT CON N ECTI ON OF
4
W I RE
Pt 1 0 0
SEN SORS
4 w ir e Pt 1 0 0
se nsor s
can
be soldered
direct ly t o t he pins of t he fly ing fem ale
connect or wit hout m ak ing use of t he
TP47 board. The 4 wires of t he Pt 100
sensors have t o be soldered as
indicat ed in t he figur e on t he left . I n
order t o use t his t y pe of probe it is
necessary t o set up t he m enu it em
“ Probe Type” , as described at page 9.
The P100 probe is recognized upon
t urning on t he inst rum ent : connect t he
probe when t he inst r um ent is sw it ched
off and t hen t urn it on.
Sensor
Direct soldering to the connector
Pt 100 4
wires
View of t he soldering side
of t he flying fem ale connect or