background image

3

Signet 3719 pH/ORP Wet-Tap

Notes on Location, Orientation and Required Clearance

• 

The 3719-1 is designed for use in pipes up to 4 in.

• 

The 3719-2 is designed for use in pipes from 6 to 12 in.

• 

Select a location that will provide suf

fi

 cient clearance to 

remove and insert the electrode.

• 

The 3719 can be mounted in any orientation, including 
horizontal and inverted. Avoid the 12 o'clock position. In the 
presence of sediment, avoid the 6 o'clock position.

• 

If inverted, use caution when removing the sensor.

 Residual 

fl

 uid may be 

present in the retraction 
housing.  Keep electrode 
connector clean and dry at 
all times.

Any angle OK

a)  3719 pH/ORP Wet-Tap
b) Customer-supplied 

Low-Pro

fi

 le Clamp-on Saddle Fitting

(ASTM sizes 2½ to 12 in.)

c) DryLoc

®

 pH or ORP Electrode (6 versions available; see 

ordering infomation on page 8.)

 (DryLoc

®

 refers to the electrode connector style)

d)  2750 or 2760 DryLoc

®

 pH/ORP Preampli

fi

 er/Sensor 

electronics

e)  Output signal options: 
  2750:
 

 

 - Digital (S

3

L)

 

 

 - 4 to 20 mA

  2760:
 

 

 - Analog mV Output- Digital (S

3

L)

All components (items a–d) are sold separately.

General

Compatible sensors:
 2756-WT 

DryLoc

®

 pH Electrode

  2756-WT-1 DryLoc pH Electrode
  2756 -WTP DryLoc plastic pH electrode
  2756-WTP-1 DryLoc plastic pH electrode
  2757-WT DryLoc ORP Electrode
  2757-WTP DryLoc plastic ORP electrode

Shipping Weight

  Wet-tap assembly ............. 1.2 kg (2.7 lbs.)
  Electrode ........................... 0.13 kg (0.3 lb)
Process Connection ............. 3-3719-11: 1½ in. NPT

3-3719-21: 2 in. NPT
3-3719-12: ISO 7/1-R1½
3-3719-22: ISO 7/1-R2

Wetted materials

Retraction Housing ............... CPVC
O-rings .................................. FKM
Electrode Body ..................... Glass or Plastic
Electrode junctions ............... Porous PTFE
Electrode sensing surface .... Glass Membrane (pH)

Platinum (ORP)

Electrode O-rings.................. FKM

Other Materials

Locking Shroud..................... PVC
Hardware .............................. 316 stainless steel

Performance

Maximum Flow Velocity ........ 3 m/s (10 ft/s)
Response Time
  pH ..................................... <5s for 95% of signal change
 ORP 

.................................. Application 

dependent

Drift ....................................... < 2 mV per week

Operating Range

  pH .....................................  0 to 14 pH
  ORP .................................. –2000 to +2000 mV
Reference Junctions ............. Porous PTFE
Electrolyte ............................. 3.5M  KCl
Elements............................... Ag/AgCl
Temp. Sensor (pH) ............... 3K Balco (3-2756-WT-1, WTP-1)

PT1000 (3-2756-WT, -WTP)

Temp. response time (

τ

) ....... 438 s

Sodium Ion Error .................. < 0.05 pH in 0.1 molar

Na+ ion at 12.8 pH

Impedance (pH) .................... <150 M

Ω

 @ 25 °C

Storage Temperature

  Wet-tap assembly ............. -15 to 120 °C (5 to 248 °F)
  Electrode ........................... 0 to 50 °C (32 to 122 °F)

Standards and Approvals

  Manufactured under ISO 9001 and 14001

       China RoHS (Go to www.gfsignet.com for details)

25

Specifi cations

System Overview

Installation

Summary of Contents for Signet 8850-3

Page 1: ...urge customers to read the specifications carefully before installing and operating this product Improper use can cause components and process liquids to be expelled at high speeds and cause serious...

Page 2: ...warranted out of box but not warranted against any damage due to process or application failures e g high temperature chemical poisoning dry out or mishandling e g broken glass damaged membrane freezi...

Page 3: ...c pH Electrode 2756 WT 1 DryLoc pH Electrode 2756 WTP DryLoc plastic pH electrode 2756 WTP 1 DryLoc plastic pH electrode 2757 WT DryLoc ORP Electrode 2757 WTP DryLoc plastic ORP electrode Shipping Wei...

Page 4: ...pe sizes below 2 5 inches by creating a flow cell with standard piping components One simple solution using a tee fitting and reducer bushings is shown in the example below Many similar configurations...

Page 5: ...operation Use an appropriate thread sealant to prevent leaks The piping system can now be safely pressurized Inspect the installation for leaks 3 4 The electrode piston is locked in position by SS lo...

Page 6: ...iate solvent that does not attack the materials of construction ORP electrode surface platinum can be gently sanded with 600 grit wet and dry silicone or carbide sandpaper jewelers rouge crocus cloth...

Page 7: ...0 06 0 0 06 0 12 0 18 0 24 0 3 55 0 45 0 36 0 27 0 18 0 09 0 0 09 0 18 0 27 0 36 0 45 4 pH w Quinhydrone 7 pH w Quinhydrone Temp 20 C 25 C 30 C 20 C 25 C 30 C ORP 268 mV 263 mV 258 mV 92 mV 86 mV 79...

Page 8: ...de pH DryLoc plastic bulb 3 K wet tap 3 2757 WT 159 000 835 Electrode ORP DryLoc bulb 10 K ID wet tap 3 2757 WTP 159 001 391 Electrode ORP DryLoc plastic bulb 10 K ID wet tap 3 2750 1 159 000 744 In l...

Reviews: