+GF+ Signet 8850-3 Operating Instructions Manual Download Page 4

4

Signet 3719 pH/ORP Wet-Tap

Installation in Pipe Sizes 2.5 to 12 Inches

• 

For reliable in-line measurements of pH and ORP, it is 
imperative to position the electrode tip into the process 
stream.

• 

Because of its compact “short stroke” design, the 3719 
requires low-pro

fi

 le 

fi

 ttings to ensure

proper positioning in pipe sizes 
DN65 to DN300 (2.5 to 12 inches).

• 

It is strongly recommended to 
use low-profi le PP clamp-on 
saddle fi ttings (customer-
supplied).

• 

Choose the 3719 version (-11 
or -21) appropriate to the size 
of the branch connection of the 
required 

fi

 tting:

 

Use -11 for sizes DN65 to DN100 
(2.5 to 4 in), and -21 for sizes 
DN150 to DN300 (6 to 12 in.)

Installation in Pipe Sizes less than 2.5 Inches

• 

It is possible to install the 3719 into pipe sizes below 
2.5 inches by creating a “

fl

 ow cell” with standard piping 

components.

• 

One simple solution, using a tee 

fi

 tting and reducer bushings, 

is shown in the example below.

• 

Many similar con

fi

 gurations are conceivable in a wide 

variety of materials, but be very careful to verify dimensional 
compatibility.

• 

Select an appropriate installation orientation to avoid the 
entrapment of air inside the 

fl

 ow cell.

• 

Contact your local Georg Fischer Sales and Support of

fi

 ce 

for assistance.

3-3719-11
pH/ORP Wet-Tap
(Fully Inserted)

2 in. Tee 
(S x S x S) 
 

1½ in. NPT  
Threads 

Flush Style Reducer 
Bushings (Spg x S) 
 

2 in. x ½ in. 

 

2 in. x ¾  in. 

 

2 in. x 1  in. 

 

2 in. x 1¼ in. 

 

2 in. x 1½ in. 

  

 

3-3719-11 
pH/ORP Wet-Tap 
 

Flush Style Reducer Bushing 
(Spg x FT)  2 in. x 1½ in. 
 

250 mm 
(9.9 in.) 

45 mm 
(1.75 in.) 

115 mm
(4.5 in.)

218 mm 
(8.57 in.) 

OR

89.4 mm

(3.5 in.)

• 

Provide 20 inches (minimum) clearance from the top of the 
pipe for electrode removal..

Installation continued

Installation continued

Summary of Contents for Signet 8850-3

Page 1: ...urge customers to read the specifications carefully before installing and operating this product Improper use can cause components and process liquids to be expelled at high speeds and cause serious...

Page 2: ...warranted out of box but not warranted against any damage due to process or application failures e g high temperature chemical poisoning dry out or mishandling e g broken glass damaged membrane freezi...

Page 3: ...c pH Electrode 2756 WT 1 DryLoc pH Electrode 2756 WTP DryLoc plastic pH electrode 2756 WTP 1 DryLoc plastic pH electrode 2757 WT DryLoc ORP Electrode 2757 WTP DryLoc plastic ORP electrode Shipping Wei...

Page 4: ...pe sizes below 2 5 inches by creating a flow cell with standard piping components One simple solution using a tee fitting and reducer bushings is shown in the example below Many similar configurations...

Page 5: ...operation Use an appropriate thread sealant to prevent leaks The piping system can now be safely pressurized Inspect the installation for leaks 3 4 The electrode piston is locked in position by SS lo...

Page 6: ...iate solvent that does not attack the materials of construction ORP electrode surface platinum can be gently sanded with 600 grit wet and dry silicone or carbide sandpaper jewelers rouge crocus cloth...

Page 7: ...0 06 0 0 06 0 12 0 18 0 24 0 3 55 0 45 0 36 0 27 0 18 0 09 0 0 09 0 18 0 27 0 36 0 45 4 pH w Quinhydrone 7 pH w Quinhydrone Temp 20 C 25 C 30 C 20 C 25 C 30 C ORP 268 mV 263 mV 258 mV 92 mV 86 mV 79...

Page 8: ...de pH DryLoc plastic bulb 3 K wet tap 3 2757 WT 159 000 835 Electrode ORP DryLoc bulb 10 K ID wet tap 3 2757 WTP 159 001 391 Electrode ORP DryLoc plastic bulb 10 K ID wet tap 3 2750 1 159 000 744 In l...

Reviews: