background image

7

Signet 3719 pH/ORP Wet-Tap

Offset in pH Electrodes

Electrode offsets occur due to:
• 

Clogged reference junction

• 

Aged or contaminated reference solution/wire

• 

A constant output near 0 mV in all buffer solutions indicates a shorted electrode 
that must be replaced.

Check offsets in a pH 7 buffer @ 25 °C. The theoretical output is 0 mV. Any deviation from 0 
mV is the pH electrode offset. The mV offset will track across the entire pH range. The slope 
is usually not affected by offset changes. 
(i.e., pH 7= +10 mV, pH 4= +187 mV); slope = 59 mV.

pH Electrode Offset pH 7 buffer @ 25 °C
 

Theoretical:   

pH 7.0 (0.0 mV)

 

New electrode: 

pH 7 ± 0.25 pH (±15 mV)

 

Reliable: 

 

pH 7 ± 0.85 pH (± 50 mV)

Electrode offsets greater than 0.85 pH (50 mV) indicate the electrode requires cleaning
or replacement. See Maintenance and Cleaning section.

Offset in ORP Electrodes

• 

ORP electrode offsets are usually caused by clogged reference junctions or by an aged or contaminated reference solution/wire.

• 

Offsets should be checked in pH 7 buffer saturated with quinhydrone @ 25 °C. The theoretical output is +86 mV.

 

Any deviation from +86 mV is the ORP electrode offset (e.g., +90 mV).

• 

Quinhydrone is the oxidizer measured by the ORP electrode and is required for calibration.

 

To measure ORP electrode offset, saturate 50 mL of pH 4 and pH 7 buffers with 

 g quinhydrone.

A new ORP electrode measures these values ±15 mV. The electrode continues to be functional until the offset from these values 
exceeds 50 mV. Electrodes with offset greater than 50 mV should be cleaned and replaced if necessary.

°C

pH

2

3

4

5

6

7

8

9

10

11

12

15

0.15

0.12

0.09

0.06

0.03

0

0.03

0.06

0.09

0.12

0.15

25

0

0

0

0

0

0

0

0

0

0

0

35

0.15

0.12

0.09

0.06

0.03

0

0.03

0.06

0.09

0.12

0.15

45

0.3

0.24

0.18

0.12

0.06

0

0.06

0.12

0.18

0.24

0.3

55

0.45

0.36

0.27

0.18

0.09

0

0.09

0.18

0.27

0.36

0.45

4 pH w/Quinhydrone

7 pH w/Quinhydrone

Temp:

20 °C

25 °C

30 °C

20 °C

25 °C

30 °C

ORP:

268 mV 263 mV 258 mV

92 mV

86 mV

79 mV

Slope in ORP electrodes

ORP slope errors are caused by contamination of the platinum electrode surface. Cleaning the electrode surface will usually restore 
proper values, response time, and stability. Many systems require both pH and ORP calibration. To conserve calibration reference 
solutions, use pH 7 and 4 buffers for pH calibration 

fi

 rst. ORP calibration can be performed with the same buffers after adding 

quinhydrone.

Slope in pH electrodes

Electrode slope is the mV output per pH unit. At 25



C the theoretical slope is 59.16 mV per pH. The graph below illustrates potential

pH error when a temperature-compensated instrument is not used.
• 

Coatings on the glass may affect sensor slopes. See Maintenance and Cleaning section.

• 

Temperature affects electrode slope. Calibrate temperature before calibrating the standard and slope.

Response Time/Stability

Response time and stability are affected by the condition of the glass surface (ORP electrode - Platinum surface), reference junction, 
and reference solution. Restoration to acceptable levels can often be accomplished by cleaning the electrode's glass surface (ORP 
electrode - platinum surface) and reference junction.
pH and ORP electrodes are similar to batteries; they age with time and usage.
The following information will help maximize electrode life:
• 

High temperatures or concentrated acids/caustics will accelerate electrode aging.

• 

Never store the electrode tip in deionized (DI) water.

• 

Never expose electrode to temperatures below –12 °C (10 °F) or allow it to dehydrate. These conditions will damage the electrode.

Troubleshooting

Theoretical mV Values @ 25 °C

pH

mV

2

+296 mV

3

+237 mV

4

+177 mV

5

+118 mV

6

+59 mV

7

0 mV

8

-59 mV

9

-118 mV

10

-177 mV

11

-237 mV

12

-296 mV

Summary of Contents for Signet 8850-3

Page 1: ...urge customers to read the specifications carefully before installing and operating this product Improper use can cause components and process liquids to be expelled at high speeds and cause serious...

Page 2: ...warranted out of box but not warranted against any damage due to process or application failures e g high temperature chemical poisoning dry out or mishandling e g broken glass damaged membrane freezi...

Page 3: ...c pH Electrode 2756 WT 1 DryLoc pH Electrode 2756 WTP DryLoc plastic pH electrode 2756 WTP 1 DryLoc plastic pH electrode 2757 WT DryLoc ORP Electrode 2757 WTP DryLoc plastic ORP electrode Shipping Wei...

Page 4: ...pe sizes below 2 5 inches by creating a flow cell with standard piping components One simple solution using a tee fitting and reducer bushings is shown in the example below Many similar configurations...

Page 5: ...operation Use an appropriate thread sealant to prevent leaks The piping system can now be safely pressurized Inspect the installation for leaks 3 4 The electrode piston is locked in position by SS lo...

Page 6: ...iate solvent that does not attack the materials of construction ORP electrode surface platinum can be gently sanded with 600 grit wet and dry silicone or carbide sandpaper jewelers rouge crocus cloth...

Page 7: ...0 06 0 0 06 0 12 0 18 0 24 0 3 55 0 45 0 36 0 27 0 18 0 09 0 0 09 0 18 0 27 0 36 0 45 4 pH w Quinhydrone 7 pH w Quinhydrone Temp 20 C 25 C 30 C 20 C 25 C 30 C ORP 268 mV 263 mV 258 mV 92 mV 86 mV 79...

Page 8: ...de pH DryLoc plastic bulb 3 K wet tap 3 2757 WT 159 000 835 Electrode ORP DryLoc bulb 10 K ID wet tap 3 2757 WTP 159 001 391 Electrode ORP DryLoc plastic bulb 10 K ID wet tap 3 2750 1 159 000 744 In l...

Reviews: