Signet 9950 Conductivity Transmitter 

English

Operating Instructions

Description

*3-9950-1P1.090*

3-9950-1P1.090   Rev 0   09/18

• 

English

• 

中文

The 9950 Conductivity Transmitter is a two channel controller that can support two sensors in 
one instrument. The sensor types supported by the 9950 are the Signet Conductivity/Resistivity 
sensors.

The 9950 includes advanced features such as derived functions, advanced multiple relay 
modes, and timer based relay functions. Derived functions allows for the control of a relay or 
current loop with the sum, delta (difference), or ratio of two measurements. Multiple relay modes 
allow up to three signals to be used for the control of a single relay, and can be any combination 
of analog and binary inputs. The timer relay modes allow a relay to be activated on a repeating 
basis from every minute to once every 30 days. Weekday timer mode allows a relay to be 
energized, on a speci

fi

 c day or days of the week at a speci

fi

 c time.

Your 9950 system may include the following modules:

• 

Direct Single Channel Conductivity Resistivity Module

• 

Dual Channel Conductivity Resistivity Module

• 

Four Channel Mechanical Relay Module (Optional)

• 

Two Mechanical and Two Solid State Relay Module (Optional)

• 

Two Mechanical Relays and Four Binary Inputs Module (Optional)

The 9950 Conductivity Transmitter is compatible with the Conductivity sensors listed in 
the table below.

• 

Conductivity/Resistivity measurement requires the conductivity resistivity module 
or Signet 2850 Conductivity/Resistivity Sensor Electronics (sold separately).

Compatibility

Cell 

Constant

¾ NPT

ISO 7/1-R3/4

Other

0.01

3-2819-1    (198 844 010)
3-2839-1V  (159 001 810)

3-2839-1VD  (159 001 811) 3-2819-S1 Sanitary 

Tri Clamp  (159 000 085)

0.1

3-2820-1    (198 844 000)
3-2840-1V  (159 001 812)

3-2840-1VD  (159 001 813)

1

3-2821-1    (198 844 001)
3-2841-1V  (159 001 814)

10

3-2822-1    (198 844 002)
3-2842-1V  (159 001 816)

3-2842-1VD  159 001 817)

20

3-2823-1    (198 844 003)

Your system may also be wired to a special order sensor with any of the following features.  

•  High Temperature and Pressure options.
•  Wetted materials (Hastelloy-C, Monel and Titanium) and sensor lengths.
•  Wet-Tap, ball valve retractable sensor for long insertion length.

Summary of Contents for Signet 9950

Page 1: ...e following modules Direct Single Channel Conductivity Resistivity Module Dual Channel Conductivity Resistivity Module Four Channel Mechanical Relay Module Optional Two Mechanical and Two Solid State...

Page 2: ...Warranty Information Product Registration Safety Information Standards and Approvals CE UL CUL WEEE RoHS Compliant Manufactured under ISO 9001 for Quality ISO 14001 for Environmental Management and OH...

Page 3: ...t arcing sufficient to damage the relays The RC Filter Kit or snubber part number 3 8050 396 is available as an accessory to reduce or eliminate these damaging effects Recommended for inductive loads...

Page 4: ...mounted using the provided quick clip mounting bracket Install plug in modules first with power disconnected Next step Wiring Step 2 panel V FREQ DATA GND V FREQ DATA GND Loop1 Loop1 Loop2 Loop2 PWR P...

Page 5: ...to eliminate fraying Insert wire tip or ferrule completely into the terminal and secure with the screw Do not allow any AC leads that may be connected to the internal relays to come in contact with l...

Page 6: ...r Terminal 3 9950 1 or 3 9950 2 Required by the instrument 12 to 32 VDC 10 regulated 0 5 A MAX DC power input and Loop current regulated voltage require the use of a UL60950 1 or UL61010 1 certified p...

Page 7: ...D TEMP SIGNAL SHLD 2819 2820 2821 Sanitary Tri clamp 2822 2839 2842 2839 2842 In case of noise interference connect the cable shield to earth ground Conductivity sensors connected to the Direct Conduc...

Page 8: ...1 high 6 0 mA at 30 VDC 4 8 mA at 24 VDC 2 4 mA at 12 VDC 1 0 mA at 5 VDC Terminals accept 12 to 24 AWG wire Strip 10 to 12 mm 0 4 to 0 5 in of insulation from wire tips and tin bare ends to eliminate...

Page 9: ...3 9950 1 MUST BE POWERED BY 12 32 VDC at 0 5 A V FREQ DATA GND V FREQ DATA GND Loop1 Loop1 Loop2 Loop2 PWR PWR CH2 CH1 DC POWER and imported par Black Red Power Supply 12 to 32 VDC Red Black PLC or R...

Page 10: ...its of Measure GPM pH sec etc Numeric Values or labels Bar Graph Channel 2 display area Numeric Values or labels Units of Measure GPM pH sec etc Password Overview The password is required to start edi...

Page 11: ...MM DD YYYY date format and or to scroll through other format options Press ENTER to save the date format selection 5 Press to move to the next selection SET DATE Press to scroll through the days mont...

Page 12: ...159 001 890 9950 Conductivity Transmitter AC Power 3 9950 1 2 159 001 891 9950 Dual Conductivity Transmitter DC Power 3 9950 2 2 159 001 892 9950 Dual Conductivity Transmitter AC Power Optional Modul...

Reviews: