background image

3

Signet 9950 Conductivity Transmiter 

 

Minimize handling of the plug-in modules to reduce the 
possibility of damage due to ESD.

 

Handle modules by the edges. 
Never touch any exposed circuitry or contacts.

 

Wear an anti-static wristband or stand on an anti-static mat, 
or keep one hand touching a properly grounded pipe or other 
piece of properly grounded metal when handling modules.

CAUTION 

Avoid Electrostatic Discharge (ESD).

NOTE: 

•  The four red Indicator LEDs on the front panel of the 9950 

show the status of relays 
1, 2, 3, & 4.  The LEDs will activate with or without a relay 
module installed.

•  Hysteresis and time delay are adjustable for each relay.

• Dry-contact relays are electromechanical switches with
  a moving contact armature. They are suitable for many
  general-purpose applications, AC or DC, including AC 
  loads up to 250 V.
• Solid-state relays are electronic switches with no moving 
  parts. They may be used with AC or DC loads, but have 
  lower current handling capability and voltage ratings than 
  Dry-contact relays. Solid-state relays will outlast Dry-
  contact relays in pulsing applications. 30 VDC/ VAC 
 maximum. 
• The four Binary Inputs can detect if an external sensor 
  or switch is open or closed. The switch state can be used 
  to control the relays in the module.  The four binary inputs 
  only control relay activation and are not considered an 
 input measurement. 

For wiring information, refer to the Relay Wiring section, 
page 8.

Switching active loads (usually inductive) can cause 
contact arcing suf

fi

 cient to damage the relays. 

The RC Filter Kit or “snubber” (part number 
3-8050.396) is available as an accessory to reduce 
or eliminate these damaging effects. Recommended 
for inductive loads greater than 50 VAC (remote 
relays, solenoids, pumps, etc.)

CAUTION

The 3-9950.393-3 module can be set to supply power 
to the external switch/sensor. For switches/sensors that 
require power set the Int/Ext switch to the Int positions. 
If the external Switches/Sensors are powered set the 
Int/Ext switch to Ext. Important for externally powered 
switches/sensors ensure the output signal is within the 
allowable range for the 3-9950.393-3.

3-9950.393-3

Keep relay wiring separate from sensor 
and signal wiring to prevent interference 
and damage to the 9950 Transmitter.

CAUTION

Int / Ext

Switch

EXT

INT

Optional modules and accessories are available for the 9950:
 

a.

  Base Unit (required)

 

b.

  Slot for Conductivity Resistivity Module

 

c.

  Slot for optional Loop Output Module

 d.

  Slot for optional Relay Module

Plug-In Modules

a.

b .

c .

Mfr. Part No.

Code

Description

3-9950.394-1

159 001 894

Single Channel Direct 
Conductivity Resistivity Module

3-9950.398-2

159 001 848

Dual Channel 4 to 20 mA 
Current Loop Output Module

Conductivity and Loop  Modules

Direct Conductivity Resistivity Module

Compatible with all GF Signet Conductivity Electrodes. 
Maximum Conductivity cable length is 30 meters (100 feet). 
For longer distances use the 3-2850 Conductivity Sensor 
Electronics. Up to two modules can be installed in a 
9950 Transmitter

Dual Channel 4 to 20 mA Module

Two additional 4 to 20 mA passive current loop outputs in a 
single module, allows up to six current loop modules in the 
9950 Transmitter.

d.

Summary of Contents for Signet 9950

Page 1: ...e following modules Direct Single Channel Conductivity Resistivity Module Dual Channel Conductivity Resistivity Module Four Channel Mechanical Relay Module Optional Two Mechanical and Two Solid State...

Page 2: ...Warranty Information Product Registration Safety Information Standards and Approvals CE UL CUL WEEE RoHS Compliant Manufactured under ISO 9001 for Quality ISO 14001 for Environmental Management and OH...

Page 3: ...t arcing sufficient to damage the relays The RC Filter Kit or snubber part number 3 8050 396 is available as an accessory to reduce or eliminate these damaging effects Recommended for inductive loads...

Page 4: ...mounted using the provided quick clip mounting bracket Install plug in modules first with power disconnected Next step Wiring Step 2 panel V FREQ DATA GND V FREQ DATA GND Loop1 Loop1 Loop2 Loop2 PWR P...

Page 5: ...to eliminate fraying Insert wire tip or ferrule completely into the terminal and secure with the screw Do not allow any AC leads that may be connected to the internal relays to come in contact with l...

Page 6: ...r Terminal 3 9950 1 or 3 9950 2 Required by the instrument 12 to 32 VDC 10 regulated 0 5 A MAX DC power input and Loop current regulated voltage require the use of a UL60950 1 or UL61010 1 certified p...

Page 7: ...D TEMP SIGNAL SHLD 2819 2820 2821 Sanitary Tri clamp 2822 2839 2842 2839 2842 In case of noise interference connect the cable shield to earth ground Conductivity sensors connected to the Direct Conduc...

Page 8: ...1 high 6 0 mA at 30 VDC 4 8 mA at 24 VDC 2 4 mA at 12 VDC 1 0 mA at 5 VDC Terminals accept 12 to 24 AWG wire Strip 10 to 12 mm 0 4 to 0 5 in of insulation from wire tips and tin bare ends to eliminate...

Page 9: ...3 9950 1 MUST BE POWERED BY 12 32 VDC at 0 5 A V FREQ DATA GND V FREQ DATA GND Loop1 Loop1 Loop2 Loop2 PWR PWR CH2 CH1 DC POWER and imported par Black Red Power Supply 12 to 32 VDC Red Black PLC or R...

Page 10: ...its of Measure GPM pH sec etc Numeric Values or labels Bar Graph Channel 2 display area Numeric Values or labels Units of Measure GPM pH sec etc Password Overview The password is required to start edi...

Page 11: ...MM DD YYYY date format and or to scroll through other format options Press ENTER to save the date format selection 5 Press to move to the next selection SET DATE Press to scroll through the days mont...

Page 12: ...159 001 890 9950 Conductivity Transmitter AC Power 3 9950 1 2 159 001 891 9950 Dual Conductivity Transmitter DC Power 3 9950 2 2 159 001 892 9950 Dual Conductivity Transmitter AC Power Optional Modul...

Reviews: