2

Signet 5500 Flow Monitor

4.2  Totalizer pulse mode (sec. 8.2)

•  External counter or accumulator inputs without internal open-collector excitation voltage and pull-up resistor.

Gnd 

PLS 

output 

Gnd 

PLS 

output 

Ÿ

Ÿ

+

-

+

-

5500  
Terminals 

28 VDC

max.

Other

instrument

Gnd

Input

PNP source input

5500  
Terminals 

External power supply

28 VDC

max.

Other

instrument

Gnd 

Input

NPN sink input

•  TTL inputs:  An external 5 VDC power supply and 10 k

Ω

 pull-up 

resistor is recommended.

•  CMOS inputs: An external 3 to 15 VDC power supply and 10 k

Ω

 

pull-up resistor is recommended.

•  12 VDC Open-collector inputs: An external 12 VDC power 

supply and 10 k

Ω

 pull-up resistor is recommended.

•  24 VDC Open-collector inputs: An external 24 VDC power 

supply and 10 k

Ω

 pull-up resistor is recommended.

3.2  Pulse output connection options for instruments without internal open-collector excitation voltage and pull-up resistor.

Gnd 

PLS 

output 

Gnd 

PLS 

output 

Ÿ

Ÿ

+

-

+

-

5500  
Terminals 

28 VDC

max.

Other

instrument

Gnd

Input

PNP source input

5500  
Terminals 

External power supply

28 VDC

max.

Other

instrument

Gnd 

Input

NPN sink input

Gnd 

AUX 

output 

Ÿ

5500 
Terminals

Signal output

Signal ground

3 - 28  

VDC 

 

External

power supply

3.  Sensor Pulse Output Connections (sec. 8.1)

3.1  Signet instrument with internal open-collector excitation 

voltage and pull-up resistor.

Technical Notes:

•  Use 2-conductor shielded twisted-pair cable for output lines up 

to 30 m (100 ft.) max.

•  Pull-up resistor NOT REQUIRED when connected to

Signet instruments.

Gnd 

PLS 

output 

Shld 

Red 

Ÿ

5500 
Terminals

Signet

Instrument

Iso. Gnd
Shld

Freq. IN*
Red

*(Open-collector sensor 
   terminals)

4.  Auxiliary Pulse Output Connections

4.1  Pulse mode (sec. 8.2) - External device control

•  The auxiliary output can be con

fi

 gured for either TTL, CMOS, 

or open-collector compatible inputs using the proper pull-up 
resistor and an external power supply.

Technical Notes:

•  Open-collector transistor, optically isolated, 5 mA max. sink,

28 VDC max. pull-up voltage.

•  Programmable pulse width, 0.10 - 999. seconds (sec. 9.2 M)
•  Selectable Active high or active low pulse logic (sec. 9.3 J).
•  Use 2-conductor shielded twisted pair cable for AUX output lines 

to 30 m (100 ft) max.

Summary of Contents for Signet 5500

Page 1: ...Double Insulated DC or AC power Contents 1 Power Connections 2 Compatible Sensor Wiring 3 Sensor Pulse Output Connections 4 Auxiliary Pulse Output Connections 5 Totalizer Reset Connections 6 4 20 mA...

Page 2: ...t PNP source input 5500 Terminals External power supply 28 VDC max Other instrument Gnd Input NPN sink input Gnd AUX output N 5500 Terminals Signal output Signal ground 3 28 VDC External power supply...

Page 3: ...wer is applied after the relay 2 setpoint is reached 6 4 to 20 mA Current Output Connections sec 8 4 4 20 mA 5500 Terminals 4 20 mA Fuse 1 8A Technical Notes 1 8 A fuse recommended customer supplied 4...

Page 4: ...nsor also displayed on totalizer To enable the totalizer function the operator must set the relay or auxiliary pulse volume setpoint to 1 0000 sec 9 2 H 9 2 L Totalizer function ideal for external cou...

Page 5: ...djustments ENTER OPTIONS sec 9 3 CALIBRATE sec 9 2 VIEW sec 9 1 2 seconds 5 seconds Press hold ENTER key for menu access Menu Displays A H Factory default displays shown A Flowrate and resettable tota...

Page 6: ...disable hysteresis feature L Sets auxiliary output pulse volume setpoint 0 0001 99999 sec 8 2 Set to 1 0000 factory default to enable totalizer function sec 8 2 M Sets auxiliary output pulse width 0...

Page 7: ...nds High 8 seconds also affects 4 to 20 mA output E Sets 4 mA output setpoint 4 mA and 20 mA setpoints are reversible F Sets 20 mA output setpoint 20 mA and 4 mA setpoints are reversible G Sets 4 mA c...

Page 8: ...ilicone rubber Weight Approximately 500 g 18 oz Display Type Microprocessor controlled air core meter movement and backlit alphanumeric 2 x 16 LCD Update rate Flow 1s totalizer 200 ms Contrast User se...

Page 9: ...96 mm 3 8 in RELAY 2 RELAY 1 ENTER 4 0 1 2 3 5 6 7 8 9 10 GPM Signet Monitor 92 x 92 mm 3 62 x 3 62 in Panel cutout Menu Parameters Display Description Range Factory Default A Flow rate and units Rese...

Page 10: ...max _Gallons E Total K Factor 0 0001 99999 60 000 F Relay 1 mode Relay 1 setpoint Low or High 0 0000 99999 Pulse 0 0001 99999 Low 10 000 GPM G Relay 1 Hysteresis Hysteresis volume 0 0000 99999 2 0000...

Page 11: ...l to D Display averaging Off 0 sec Low 4 sec High 8 sec Off E 4 mA setpoint Unlimited 0 0000 GPM F 20 mA setpoint Unlimited 100 00 GPM G 4 mA adjust 3 0 to 5 0 mA 4 00 mA H 20 mA adjust 19 to 21 mA 20...

Page 12: ...mall Reduce auxiliary output pulse width setting sec 9 2 M or increase pulse volume setting sec 9 2 L Must be between 0 0001 99999 Relay 1 Relay 2 or Auxiliary output AUX pulse volume setpoint must be...