Signet 5500 Flow Monitor

9.1 VIEW 

Menu 

Flow & resettable total 

Permanent total 

Last calibration 

Loop output 

Press keys in sequence, 

A. 

B. 

B. 

C. 

A. 

OPTIONS = 

OPTIONS = 

To reset  

total press: 

ENTER 

ENTER 

To reset  

total press: 

Range 

(display shows:          ) 

OR 

Relay 1 

Relay 2 

Aux output 

quick  

press 

To exit without changes: 

(sec. 9.3I) 

(sec. 9.3I) 

A. 

B. 

C. 

D. 

E. 

F. 

G. 

H. 

Choose: Change: 

Save: 

H↓ 

A↑ 

3

Menus:
• VIEW menu

 (sec. 9.1): The VIEW menu is displayed during 

standard operation. The operator can navigate freely through the 
menu by pressing either UP or DOWN arrow keys. The VIEW 
menu also offers a resettable totalizer feature that allows the 
operator to reset the displayed total at the press of a button. 

• CALIBRATE Menu

 (sec. 9.2): The CALIBRATE menu contains 

all critical display setup and output parameters. A simple security 
code feature prevents unauthorized tampering. The operator is 
required to enter a simple access code for menu access. The 
same code also unlocks OPTIONS menus.

• OPTIONS Menu

 (sec. 9.3): The OPTIONS menu contains setup 

and display features that are seldom accessed for minor display 
or output adjustments.

ENTER

OPTIONS 

(sec. 9.3) 

CALIBRATE 

(sec. 9.2) 

VIEW

(sec. 9.1)

2

seconds

5

seconds

Press & hold ENTER key 
for menu 
access:

Menu Displays A - H: 

(Factory default displays shown)

A.  Flowrate and resettable total display: Can be 

reset by the front panel keypad or by remote 
reset switch (sec. 5). A simple access code 
feature prevents accidental total reset, 
when enabled (sec. 9.3 I). External reset 
switch inputs override security feature, when 
enabled.

B.  Permanent total display: Records volumetric 

fl

 ow total for the life of the product and is 

non-resettable.

C.  Loop output display: Shows the current 

output level at the rear terminals.

D.  Flow Range display: Shows the programmed 

min and max meter range (sec. 9.2 B)

E.  Relay 1 display: This display shows the 

programmed operation mode and setpoint 
for relay 1 (sec. 9.2 F-I)

F.  Relay 2 display: This display shows the 

programmed operation mode and setpoint 
for relay 2 (sec. 9.2 F-I).

G.  Aux output display: This display shows the 

programmed operation mode and setpoint 
for the auxiliary output (sec. 9.2 J-M).

H.  Last calibration: This display shows a user- 

de

fi

 ned setup date for maintenance records. 

This feature is not an internal timer or 
calendar.

9. Menu 

Functions

To access either CALIBRATE or OPTIONS menus, press and hold 
the ENTER key as illustrated below:

Summary of Contents for Signet 5500

Page 1: ...Double Insulated DC or AC power Contents 1 Power Connections 2 Compatible Sensor Wiring 3 Sensor Pulse Output Connections 4 Auxiliary Pulse Output Connections 5 Totalizer Reset Connections 6 4 20 mA...

Page 2: ...t PNP source input 5500 Terminals External power supply 28 VDC max Other instrument Gnd Input NPN sink input Gnd AUX output N 5500 Terminals Signal output Signal ground 3 28 VDC External power supply...

Page 3: ...wer is applied after the relay 2 setpoint is reached 6 4 to 20 mA Current Output Connections sec 8 4 4 20 mA 5500 Terminals 4 20 mA Fuse 1 8A Technical Notes 1 8 A fuse recommended customer supplied 4...

Page 4: ...nsor also displayed on totalizer To enable the totalizer function the operator must set the relay or auxiliary pulse volume setpoint to 1 0000 sec 9 2 H 9 2 L Totalizer function ideal for external cou...

Page 5: ...djustments ENTER OPTIONS sec 9 3 CALIBRATE sec 9 2 VIEW sec 9 1 2 seconds 5 seconds Press hold ENTER key for menu access Menu Displays A H Factory default displays shown A Flowrate and resettable tota...

Page 6: ...disable hysteresis feature L Sets auxiliary output pulse volume setpoint 0 0001 99999 sec 8 2 Set to 1 0000 factory default to enable totalizer function sec 8 2 M Sets auxiliary output pulse width 0...

Page 7: ...nds High 8 seconds also affects 4 to 20 mA output E Sets 4 mA output setpoint 4 mA and 20 mA setpoints are reversible F Sets 20 mA output setpoint 20 mA and 4 mA setpoints are reversible G Sets 4 mA c...

Page 8: ...ilicone rubber Weight Approximately 500 g 18 oz Display Type Microprocessor controlled air core meter movement and backlit alphanumeric 2 x 16 LCD Update rate Flow 1s totalizer 200 ms Contrast User se...

Page 9: ...96 mm 3 8 in RELAY 2 RELAY 1 ENTER 4 0 1 2 3 5 6 7 8 9 10 GPM Signet Monitor 92 x 92 mm 3 62 x 3 62 in Panel cutout Menu Parameters Display Description Range Factory Default A Flow rate and units Rese...

Page 10: ...max _Gallons E Total K Factor 0 0001 99999 60 000 F Relay 1 mode Relay 1 setpoint Low or High 0 0000 99999 Pulse 0 0001 99999 Low 10 000 GPM G Relay 1 Hysteresis Hysteresis volume 0 0000 99999 2 0000...

Page 11: ...l to D Display averaging Off 0 sec Low 4 sec High 8 sec Off E 4 mA setpoint Unlimited 0 0000 GPM F 20 mA setpoint Unlimited 100 00 GPM G 4 mA adjust 3 0 to 5 0 mA 4 00 mA H 20 mA adjust 19 to 21 mA 20...

Page 12: ...mall Reduce auxiliary output pulse width setting sec 9 2 M or increase pulse volume setting sec 9 2 L Must be between 0 0001 99999 Relay 1 Relay 2 or Auxiliary output AUX pulse volume setpoint must be...