background image

4

Signet 5500 Flow Monitor

A.  LOW alarm mode  
The output is active (relay energized) when the flow drops below the  
setpoint, and is inactive (relay de-energized) when the flow rises 
above the setpoint plus hysteresis (sec. 9.2 F-G, 9.2 J-K). 
 
• Auxiliary output: ideal for external device 
  inputs (sec. 4) 
• Relay outputs: ideal for external warning  
  lamps, sirens, etc. (sec. 7) 

B.  HIGH alarm mode  
The output is active (relay energized) when the flow rises above the  
setpoint and is inactive (relay de-engergized) when the flow falls 
below the setpoint plus hysteresis (sec. 9.2 F-G, 9.2 J-K). 
 
• Auxiliary output: ideal for external device  
  inputs (sec. 4) 
• Relay outputs: ideal for external warning  
  lamps, sirens, etc. (sec. 7) 
 
 

Flo

Flow

C.  PULSE mode  
The output is active (relay energized) for the programmed pulse width each time the programmed pulse volume is measured by the flow 
sensor  (sec 9.2 H-I, 9.2 L-M).   
 
• Auxiliary output: ideal for external device control (sec. 4) 
• Relay outputs: ideal for metering pump or solenoid valve control (sec. 7) 

FLOW 

volume 1 

= Programmed pulse  
  volume (e.g. 12.5 Gal.) 

= Programmed pulse  
  width (relay energize time) 

= Output active 
 (relay 

energized) 

= Output inactive  
 (relay 

de-energized) 

volume 

volume 2 

volume 3 

volume 4 

volume 5 

volume 6 

Flow 

 Sensor 

Low setpoint= 

Adjustable hysteresis= 

Output active (relay energized)= 

Output inactive (relay de-energized)= 

High setpoint= 

Adjustable hysteresis= 

Output active (relay energized)= 

Output inactive (relay de-energized)= 

Totalizer function 
The output emits an adjustable signal pulse width for each whole engineeing unit measured by the flow sensor (also displayed on totalizer).  
To enable the totalizer function, the operator must set the relay or auxiliary pulse volume setpoint to 1.0000 (sec. 9.2 H, 9.2 L). 
 
• Totalizer function: ideal for external counter or accumulator use (sec 4.2) 

adjustable 

pulse width 

VIEW menu  

totalizer display 

8.3  Totalizer reset (sec. 5): 

This feature allows the operator to reset the totalizer with a dry switch contact closure at any time. The 

external reset switch overrides software security feature (sec. 5).

8.4  4 to 20 mA output (sec. 6):

 The 4 to 20 mA output is internally powered by the instrument. 4 mA and 20 mA setpoints are fully 

adjustable within the programmed 

fl

 ow range (sec. 11 B). The output can also be reverse spanned.

8.5  Relay 1 and 2 outputs (sec 7): 

Relay outputs 1 and 2 can be independently con

fi

 gured as Low alarm, High alarm, Pulse or Totalizer 

output. Refer to diagrams A - C above for operation details.

8. Output 

Functions

This section gives a detailed description of all 5500 output functions.

8.1  Sensor pulse output (sec. 3):

 This output emits an open-collector frequency equal to the sensor input. The output connects directly 

to a second Signet instrument for expanded process control with a single sensor input (sec. 3.1). The output is an open-collector type that 
requires an external pull up resistor and power supply for other instrument inputs (sec. 3.2).

8.2  Auxiliary pulse output (sec. 4): 

The auxiliary output can be con

fi

 gured as Low alarm, High alarm, Pulse or Totalizer output. Output 

logic can be selected for either active low (factory default) or active high operation (sec. 9.3 J). The auxiliary output is an open-collector 
type that requires an external pull-up resistor and power supply to operate. Refer to diagrams A - C below for operation details. 

Summary of Contents for Signet 5500

Page 1: ...Double Insulated DC or AC power Contents 1 Power Connections 2 Compatible Sensor Wiring 3 Sensor Pulse Output Connections 4 Auxiliary Pulse Output Connections 5 Totalizer Reset Connections 6 4 20 mA...

Page 2: ...t PNP source input 5500 Terminals External power supply 28 VDC max Other instrument Gnd Input NPN sink input Gnd AUX output N 5500 Terminals Signal output Signal ground 3 28 VDC External power supply...

Page 3: ...wer is applied after the relay 2 setpoint is reached 6 4 to 20 mA Current Output Connections sec 8 4 4 20 mA 5500 Terminals 4 20 mA Fuse 1 8A Technical Notes 1 8 A fuse recommended customer supplied 4...

Page 4: ...nsor also displayed on totalizer To enable the totalizer function the operator must set the relay or auxiliary pulse volume setpoint to 1 0000 sec 9 2 H 9 2 L Totalizer function ideal for external cou...

Page 5: ...djustments ENTER OPTIONS sec 9 3 CALIBRATE sec 9 2 VIEW sec 9 1 2 seconds 5 seconds Press hold ENTER key for menu access Menu Displays A H Factory default displays shown A Flowrate and resettable tota...

Page 6: ...disable hysteresis feature L Sets auxiliary output pulse volume setpoint 0 0001 99999 sec 8 2 Set to 1 0000 factory default to enable totalizer function sec 8 2 M Sets auxiliary output pulse width 0...

Page 7: ...nds High 8 seconds also affects 4 to 20 mA output E Sets 4 mA output setpoint 4 mA and 20 mA setpoints are reversible F Sets 20 mA output setpoint 20 mA and 4 mA setpoints are reversible G Sets 4 mA c...

Page 8: ...ilicone rubber Weight Approximately 500 g 18 oz Display Type Microprocessor controlled air core meter movement and backlit alphanumeric 2 x 16 LCD Update rate Flow 1s totalizer 200 ms Contrast User se...

Page 9: ...96 mm 3 8 in RELAY 2 RELAY 1 ENTER 4 0 1 2 3 5 6 7 8 9 10 GPM Signet Monitor 92 x 92 mm 3 62 x 3 62 in Panel cutout Menu Parameters Display Description Range Factory Default A Flow rate and units Rese...

Page 10: ...max _Gallons E Total K Factor 0 0001 99999 60 000 F Relay 1 mode Relay 1 setpoint Low or High 0 0000 99999 Pulse 0 0001 99999 Low 10 000 GPM G Relay 1 Hysteresis Hysteresis volume 0 0000 99999 2 0000...

Page 11: ...l to D Display averaging Off 0 sec Low 4 sec High 8 sec Off E 4 mA setpoint Unlimited 0 0000 GPM F 20 mA setpoint Unlimited 100 00 GPM G 4 mA adjust 3 0 to 5 0 mA 4 00 mA H 20 mA adjust 19 to 21 mA 20...

Page 12: ...mall Reduce auxiliary output pulse width setting sec 9 2 M or increase pulse volume setting sec 9 2 L Must be between 0 0001 99999 Relay 1 Relay 2 or Auxiliary output AUX pulse volume setpoint must be...

Reviews: