Datalogger Operation

The 0250 can serve as a fi eld data logger that downloads data directly into a *.csv (Comma Separated Value) fi le.
Note: The 0250 does NOT have internal memory to store data. It must be connected to a computer to use the datalog function.

•  Select the sensor type, click on "Read".
• Click 

Monitor/Verify Sensor

 to open a datalog setup window.

•  Enter the Logging Interval. This value represents the time between log records.
  The minimum interval is 1 second, and the maximum interval is 86400 seconds (24 hours). Example: If the Log Interval is set to

60 seconds, the 0250 will record the temperature once every minute.

•  The 0250 saves data fi les in *.csv format. The maximum number of records allowed for this type of fi le is 65535.
  If the logging interval is 60 seconds, the largest fi le possible is 65535 records, or 1092 hours of continuous recorded data.
•  To initiate a log session click 

Begin Monitor

 or 

Begin Log to File

.

Begin Monitor

•  This function starts monitoring the sensor output and displays values on a graph, but does not store the data permanently. During 

the monitoring session the 

Begin Log to File

 and 

Save Log

 functions are disabled, while the 

Pause

 and 

End Log

 functions are 

enabled. 

Save Log

 enables a monitoring session to be saved immediately after 

End log

 is clicked.

Begin Log to File

•  This function prompts the user to create a fi le on a computer drive before it starts monitoring. The system then begins recording the 

sensor output and display values on a graph. During a Log-to-fi le session the 

Begin Monitor

 and 

Save Log

 functions are disabled, 

while the 

Pause

 and 

End Log

 functions are enabled.

Sample of data fi le downloaded to Microsoft Excel

Summary of Contents for Signet 3-0250

Page 1: ...100 C 4 to 212 F Relative Humidity 0 to 90 non condensing Operating Temperature 15 to 55 C 5 to 131 F module only 3 To run the program from the CD Right click the START button and select EXPLORE Navig...

Page 2: ...setup capability of the 0250 is more complex with the magmeters than with other sensors Read these instructions carefully before attempting to modify a Magmeter CAUTION Do not connect power directly...

Page 3: ...s Set the time the Magmeter will use as the averaging period Example With averaging at 14 seconds each display is an average of the previous 14 seconds input Use higher averaging times to smooth the d...

Page 4: ...exit the program If the settings will be used again click SAVE Name the file and save it on the computer hard drive Read Read and display the existing settings from the Magmeter Write Write from the d...

Page 5: ...atic The best solution to these problems is to correct any piping deficiency that causes the instability This may involve longer straight runs upstream reducing the pipe size to maintain a full pipe a...

Page 6: ...lay into the Magmeter Click Save to save the settings from the display to a computer file 5 To use a saved file Click Load Navigate to the saved file Select Open Click Write Click Read to confirm Sign...

Page 7: ...to the saved file Select Open Click Write Click Read to confirm Signet 2750 pH Sensor Setup 2750 Wiring 2750 Operation 1 Select the engineering units for the application pH or ORP 2 Set the pH or ORP...

Page 8: ...0 seconds the 0250 will record the temperature once every minute The 0250 saves data files in csv format The maximum number of records allowed for this type of file is 65535 If the logging interval is...

Reviews: