2

Signet 0250 Setup Tool

Connect the Magmeter to the computer:

•  Connect the Magmeter to the 3-0250 as illustrated.
•  For 3-2551-11 models: The frequency/digital jumper MUST be in the Digital position.
•  Connect the isolated 24 VDC power source to 4-20 mA model Magmeters.
•  Connect the Signal Converter to any open USB port on the computer.

CAUTION!

TURN 24 VDC POWER OFF WHEN WIRING THE MAGMETER.

System connections
Signet 2552 Metal Magmeter

System connections
Signet 2551 Blind Magmeter

2551 and 2552 Magmeters

The setup capability of the 0250 is more complex with the magmeters than with other sensors. Read these instructions carefully before 
attempting to modify a Magmeter.

CAUTION!

Do not connect power

directly to this device.

White

Blue

Brown

Black

Isolated
24 VDC

+

-

White

Blue

Wire Nut

To  -VDC

*Wire 1

*Wire 2

*Wire 3

* Additional wire required

Signet 2552
4-20 mA Out

Data

+5 VDC

Ground

White

Brown

Black

Blue

Signet 2552
Frequency/S

3

L Out

1

2

3

4

2551 Blind Magmeter
Frequency/S

3

L Out

JP2

FREQUENCY OUT
DIGITAL (S  L) OUT

3

Freq/Digital jumper
must be in Digital position

GND
Data
5V out+

USB

1

2

3

4

GND

Data

5V out+

Isolated
24 VDC

+

-

2551 Blind Magmeter
4-20 mA Out

USB

CAUTION!

Do not connect power

directly to this device.

Summary of Contents for Signet 3-0250

Page 1: ...100 C 4 to 212 F Relative Humidity 0 to 90 non condensing Operating Temperature 15 to 55 C 5 to 131 F module only 3 To run the program from the CD Right click the START button and select EXPLORE Navig...

Page 2: ...setup capability of the 0250 is more complex with the magmeters than with other sensors Read these instructions carefully before attempting to modify a Magmeter CAUTION Do not connect power directly...

Page 3: ...s Set the time the Magmeter will use as the averaging period Example With averaging at 14 seconds each display is an average of the previous 14 seconds input Use higher averaging times to smooth the d...

Page 4: ...exit the program If the settings will be used again click SAVE Name the file and save it on the computer hard drive Read Read and display the existing settings from the Magmeter Write Write from the d...

Page 5: ...atic The best solution to these problems is to correct any piping deficiency that causes the instability This may involve longer straight runs upstream reducing the pipe size to maintain a full pipe a...

Page 6: ...lay into the Magmeter Click Save to save the settings from the display to a computer file 5 To use a saved file Click Load Navigate to the saved file Select Open Click Write Click Read to confirm Sign...

Page 7: ...to the saved file Select Open Click Write Click Read to confirm Signet 2750 pH Sensor Setup 2750 Wiring 2750 Operation 1 Select the engineering units for the application pH or ORP 2 Set the pH or ORP...

Page 8: ...0 seconds the 0250 will record the temperature once every minute The 0250 saves data files in csv format The maximum number of records allowed for this type of file is 65535 If the logging interval is...

Reviews: