background image

10

GF 263X Series Chlorine Electrodes 

Overview

pH Compensation for Free Chlorine

Amperometric free chlorine sensors measure only hypochlorous 
acid. As noted in the text above and in Figure 3, the ratio of 
hypochlorous acid and hypochlorite is pH dependent. 
The GF Free Chlorine Panel assemblies comes complete with 
a pH electrode to ensure accurate free chlorine measurement if 
the pH changes. 

5.0  5.5     6.0     6.5     7.0    7.5     8.0     8.5     9.0     9.5  10.0

100

90

80

70

60

50

40

30

20

10

0

HOCl
OCl

Ø

%  free chlorine

pH  at 25 °C

Chlorine Measurement by Amperometric Sensors

GF chlorine sensors are membrane-covered amperometric 2-electrode sensors. A gold or platinum cathode acts as the working 
electrode with a silver halide acting as the counter electrode. Depending on the species to be analyzed, a polarization voltage is 
applied between the two electrodes. When placed into service, the chlorine species of interest diffuses across the membrane and 
is reduced at the cathode surface. For the case of total chlorine, the analyte reacts with the 

fi

 ll solution to produce an intermediate, 

which is subsequently reduced at the cathode surface. At the same time, the silver anode is oxidized to form a silver halide. The 
current generated at the cathode is proportional to the rate of diffusion through the membrane and the concentration of chlorine in the 
sample. The current from the cathode to the anode is conditioned, digitized and transmitted by the associated electronics.

Chlorine in Water

Various forms of chlorine are used to disinfect water. Each form of chlorine has bene

fi

 ts and limitations which help determine the 

speci

fi

 c application. The predominant categories used in disinfection are Free Chlorine, Total Chlorine and Chlorine Dioxide. Free 

Chlorine is the sum of chlorine gas (Cl

2

), hypochlorous acid (HOCl) and hypochlorite (OCl

-

). Above pH 4.0 all of the molecular 

chlorine is converted to HOCl and OCl

-

. Hypochlorous acid is a more potent disinfectant than hypochlorite and exists in a pH 

dependent equilibrium as shown in Figure 3.
Free chlorine also combines with naturally occurring or human-introduced nitrogen compounds in the water to form chloramines, also 
known as combined chlorine. Treatment operators introduce ammonia into the water to form monochloramine (NH

2

Cl), dichloramine 

(NHCl

2

) and trichloramine (NCl

3

). Chloramines are a less effective disinfectant but have a longer residence time than the free chlorine 

species. Total chlorine is the sum of free chlorine (Cl

2

, HOCl and OCl

-

) and combined chlorine (NH

2

Cl, NHCl

2

, NCl

3

).

Automatic pH Compensation and Free Chlorine

In many applications, the process pH does not signi

fi

 cantly 

fl

 uctuate and only a chlorine sensor and instrument are 

necessary for accurate chlorine measurement.
When the pH varies the free chlorine concentration cannot 
accurately determined without the use of automatic pH 
compensation.

The addition of the 3-2724-00 (159 001 545) pH electrode along 
with its 3-2751-7 (159 001 957) sensor electronics to the system 
makes pH compensation extremely easy and automatic even 
with wide 

fl

 uctuations or high pH.

See Figure 4 for pH variation recommendations.

Chlorine dioxide is not pH dependant and does not require a pH 
electrode or electronics.

                6.5          7.0          7.5            8.0           8.5           9.0

0

±0.3

±0.1

±0.2

=

  pH variation

Sample pH 

   Automatic pH compensation recommended
   in ranges within shaded area

Figure 3

Figure 4

Summary of Contents for 3-2630-X

Page 1: ...nted DryLoc connector provides quick assembly and a secure connection Gold plated contacts and an O ring seal ensure a waterproof and reliable interconnect to the 2650 Amperometric Electronics Integrated temperature element for automatic temperature compensation Separate drive electronics 2650 allow easy electrode replacement without running new cable English 3 2630 090 3 2630 090 Rev 15 02 21 GF ...

Page 2: ...ctrocution Electrostatic Discharge ESD Alerts user to risk of potential damage to product by ESD Personal Protective Equipment PPE Always utilize the most appropriate PPE during installation and service of GF products Pressurized System Warning Sensor may be under pressure take caution to vent system prior to installation or removal Failure to do so may result in equipment damage and or serious in...

Page 3: ...fsignet com for details Declaration of Conformity according to FCC Part 15 This device complies with Part 15 of the FCC rules Operation is subject to the following two conditions 1 This device may not cause harmful interference and 2 This device must accept any interference received including interference that may cause undesired operation Specifications General Polarization Source GF 2650 Amperom...

Page 4: ...r a sink is recommended 1 Remove the protective bottle from the end of the electrode 2 Remove the membrane cap from the front of the sensor Note When new sensors are shipped the membrane cap is not tightened to the sensor 3 Fill supplied syringe with electrolyte solution 4 Place the electrode on a level surface 5 Insert syringe needle fully into one of the eight electrode holes while injecting wit...

Page 5: ...e operating pressure must be less than 0 48 bar 7 psi Higher pressures will damage the electrode The electrode should not be used in water containing surfactants oils organic chlorine or stabilizers such as cyanuric acid Part Number Chlorine Range Chlorine Type 3 2630 1 0 02 to 2 ppm mg L Free Chlorine 3 2630 2 0 05 to 5 ppm mg L 3 2630 3 0 1 to 20 ppm mg L 3 2632 1 0 02 to 2 ppm mg L Chlorine Dio...

Page 6: ...injecting with electrolyte solution Slowly injecting the electrolyte solution into the sensor to avoid introducing air bubbles The electrode holds approximately 14 milliliters of solution Slowly fill until solution begins to flow out of holes Do not allow the solution to run down the electrode and wet the electrical contacts in the DryLoc connector 6 Slowly screw on the membrane cap finger tight D...

Page 7: ...ng bottle with tap water added Storage periods more than 1 week Remove the membrane cap and internal electrolyte solution Rinse the sensor internal chamber with DI water or cold tap water drain and allow to dry Place the membrane cap back onto the sensor INSTALL LOOSE DO NOT COMPLETELY TIGHTEN THE CAP WHEN STORED DRY THE MEMBRANE CAP MUST BE STORED RELAXED AND UNSTRESSED Store sensor DRY in the sh...

Page 8: ...teps 11 17 11 Fill beaker with a 12 mm inch of the appropriate solution 12 Position or suspend the sensor 6 mm to 12 mm in to in above the appropriate solution DO NOT SUBMERGE THE SENSOR See Figure 2 13 Apply power to the system 14 Monitor the nA of the sensor press the down arrow three times on the 9950 3 Chlorine Controller The nA reading should start to rise Response time and nA reading will de...

Page 9: ... fitting Insert until fully seated Do Not Use Lubricant or Sealing Tape on Threads Do Not Overtighten Do Not Use Tools pipe tee When using the chlorine panel flow cell it is not necessary to lubricate this O ring Mounting Position PVC Tee Best Flow Direction Good 45º or or 45º Mount the flow cell where the sensor will be easily accessible To avoid air bubble entrapment do not mount with downward f...

Page 10: ...tations which help determine the specific application The predominant categories used in disinfection are Free Chlorine Total Chlorine and Chlorine Dioxide Free Chlorine is the sum of chlorine gas Cl2 hypochlorous acid HOCl and hypochlorite OCl Above pH 4 0 all of the molecular chlorine is converted to HOCl and OCl Hypochlorous acid is a more potent disinfectant than hypochlorite and exists in a p...

Page 11: ...present in water See technical data Surfactants in water Remove surfactants and replace cap Membrane cap coated Clean or replace membrane cap Membrane cap loose Tighten or replace membrane cap pH outside working range See Specifications on page 3 Sensor output very low Sensor conditioning time too short Condition for 4 hours minimum prior to initial calibration Chlorine content too low Add chlorin...

Page 12: ...de 0 1 to 20 ppm mg L 3 2632 1 159 001 767 Chlorine Dioxide electrode 0 02 to 2 ppm mg L Accessories and Replacement Parts Mfr Part No Code Description 3 2630 391 159 001 674 Free Chlorine electrolyte 30 mL 3 2632 391 159 310 160 Chlorine Dioxide electrolyte solution 30 mL 3 2630 394 159 310 164 Free Chlorine and Chlorine Dioxide Replacement PTFE membrane 1 3 2630 398 159 310 166 Free Chlorine Sen...

Reviews: