Premature halting of the aeration process will allow the
outer door to be opened with a residual level of H
2
O
2
in the
machine which is higher than the minimum level defined for
normal operation. This therefore represents a risk of
inhalation for the operator when opening the outer door of
the biodecontamination airlock.
−
If the "NO" button is pressed, the cycle continues.
−
The following view appears during production:
−
If the "YES" button is pressed, the cycle is halted and the isolator is no longer sterile.
−
If the "NO" button is pressed, the cycle continues.
8.4. EMERGENCY AERATION
In the event of a problem during the biodecontamination phase, the operator can launch
emergency aeration by pressing the red button next to the "EMERGENCY AERATION"
touch screen.