7.6. HANDLING
Before each sterilization cycle, Getinge-La Calhène recommends checking the integrity of
the isolator gloves. The method consists of:
−
conducting a test by raising the oxygen level to assess glove performance (
GLT
testing device – operator manual NT 3020/1
)
If the result of the test is "not correct", the glove must be replaced.
The procedure for replacing a glove without breach of containment is as follows:
Installation of gloves for an isolator under positive pressure
−
Pull the glove
(2)
to be replaced inside out into the sleeve
(1)
.
−
Remove the tightening ring
(3)
.
−
Position the new glove
(5)
parallel to the cuff ring
(4)
aligning the axis of the thumb
with the seal of the sleeve.
−
Install the new glove
(5)
on the cuff ring
(4)
up from the bead
(6)
of the glove to be
replaced.
−
Place the tightening ring
(3)
around the glove behind the bead
(6)
.
−
Position the bead
(6)
of the glove and the tightening ring
(3)
on the end of the cuff
ring groove
(4)
.
−
Remove the glove
(2)
from the sleeve
(1)
.
−
Extract the glove
(2)
.
1
2
3
4
5
6
1
2
3
4
6
Summary of Contents for ISOTEST
Page 1: ...ISOTEST USER MANUAL...
Page 62: ...NTA 3015 78 C Appendix 1 DEX Page 1 1...
Page 63: ...NTA 3015 78 C Appendix 2 DEX Page 1 11...
Page 64: ...NTA 3015 78 C Appendix 2 DEX Page 2 11...
Page 65: ...NTA 3015 78 C Appendix 2 DEX Page 3 11...
Page 66: ...NTA 3015 78 C Appendix 2 DEX Page 4 11...
Page 67: ...NTA 3015 78 C Appendix 2 DEX Page 5 11...
Page 68: ...NTA 3015 78 C Appendix 2 DEX Page 6 11...
Page 69: ...NTA 3015 78 C Appendix 2 DEX Page 7 11...
Page 70: ...NTA 3015 78 C Appendix 2 DEX Page 8 11...
Page 71: ...NTA 3015 78 C Appendix 2 DEX Page 9 11...
Page 72: ...NTA 3015 78 C Appendix 2 DEX Page 10 11...