14
Attention!
If the equipment is not fully charged, data
may be lost.
Always fully charge all components of
the equipment outside of the
potentially explosive atmosphere
before use.
Immediately charge the equipment
outside the potentially explosive
atmosphere if the charge indicator on
the data collector display is in the red
zone.
If the battery charge is too low the COM Box will
switch off automatically. If it is then switched on
again without prior recharging the battery may be
exhausted completely. The battery may be
damaged and data may be lost.
To avoid this allow the battery to charge for at
least 30 minutes before switching the COM box
on again.
Before using the equipment for the first time, you
must charge the batteries in the Com box and the
data collector outside the potentially explosive
atmosphere.
If the temperature of the battery is too high
or too low the charging process of the data
collector and the Com Box is stopped
automatically to prevent the batteries from
overloading.
The max. temperature is 35 °C (95 °F),
the min. temperature 0 °C (32 °F).
The charging process will continue
automatically as soon as the temperature
of the battery is again within the
admissible temperature limits.
Charge the data collector battery outside of the
potentially explosive atmosphere as described in
the data collector installation & operating
manual.
To connect the Com box to mains electricity with
different mains sockets, you can connect various
adapters to the power supply unit. Proceed as
follows to change the adapter:
Press the catch (15) on the adapter.
Disconnect the adapter (14) from the power
supply unit (13).
Push the required adapter onto the power
supply unit until it clicks into place.
Make sure that the USB connector is securely
inserted in the USB port (16).
The drawing shows the power supply unit for the
Com Box (stock code number 393226). The power
supply unit (stock code number 393080) can also
be used.
DANGER
Sparks generated when charging the
equipment can cause an explosion.
Recharge the batteries of the Com box
and the data collector only outside the
explosion-risk area.