C. Gerhardt GmbH & Co. KG
30
VAPODEST 20
Service and cleaning
Condensate that might escape is collected in the drip tray. Plea-
se clean the drip tray regularly.
Regularly check the consistance of the tubing and tubing connec-
tions.
Glass parts should be cleaned before long periods of non-usage
(i.e. holidays). This way obstructions caused by crystalline depo-
sits are avoided.
The following program should be run:
Addition NaOH:
0 s
Reaction time:
0 s
Distillation time:
420 s
Steam capacity:
100 %
It is recommended that the NaOH line is flushed daily so that no
NaOH is left in the chemical lines overnight.
Place an digestion tube with 70 ml H
2
O and an Erlenmeyer flask
into position, and start the program.
Trouble shooting
The excess current switch cuts off in case of excessive current
consumption. Switch on the excess current switch again. If this
problem reoccurs, call service!
In case of a breakdown or a failure of your distillation system
Vapodest 20, please contact your local dealer or:
C. Gerhardt GmbH & Co. KG
Cäsariusstr. 97
53639 Königswinter
E-Mail:
[email protected]
www.gerhardt.de
Service
Excess current switch