3.2
The technical state of Spirostik Complete and construction
Possible danger to life.
Reason: Electric shock. Cross-contamination.
Misdiagnosis caused by measurement error.
Therefore:
•
Do not overstress Spirostik Complete!
Use with care!
•
Do not modify or use Spirostik Complete or total
system contrary to the respective
manufacturer's specification!
•
Under no circumstances use or connect any
devices, systems, equipment and other products
that are not part of the total system! This also
includes only peripherals components approved
by the manufacturer!
•
Ensure that in the patient environment a distance
of 1.5 meters to the patient there are no
accessible electrical parts (interfaces, plugs, etc.)
that are not isolated from the mains with an
isolation voltage of 4 kV!
•
Ensure that the users or third party persons do
not touch the patient and any conductive
connections or parts of the device that are
located outside the patient environment at the
same time!
•
Ensure that Spirostik Complete cannot fall
down! Otherwise, the functionality must be
checked properly before putting it into
operation!