Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Version: 8 | Release date: 16. August 2021
Page 89
12.2
Installation and Operating Conditions
The following conditions supplement chap. 2.2 "Intended Use"
and must be observed to maintain the properties of the Diffustik
resp. the variant PFTstik guaranteed by the manufacturer and for
the safety of the patient and user / operator.
Storage / Transport:
min.
max.
Temperature:
-10 °C
+50 °C
Relative air humidity:
0 %
95 %
non-condensing
Operation:
min.
max.
Environmental temperature:
+10 °C
+35 °C
Avoid extreme
fluctuations in
temperature!
Relative air humidity:
20 %
95 %
non-condensing
(at least. 30 % for synthetic flooring)
Atmospheric pressure:
800 hPa
1080 hPa
Setup:
mobile
Space requirement:
(complete with medical cart)
(L) 900 mm x (W) 900 mm x (H) 1900 mm
Required floor load capacity: 150 kg/m²
Flooring:
stable; even; conductive
Environment / room:
Closed, clinical area
air-conditioned
not an explosive or flammable environment!
not near an MRI!
well ventilated
Operating Mode:
Continuous operation