Instructions for Use
Diffustik / Variant PFTstik
Serial no.: xx|8|401|yyy and 2401xxxxx
Version: 8 | Release date: 16. August 2021
Page 19
2.2
Intended Use
Diffustik resp. the variant PFTstik can be operated independently
as a non-stationary device or in combination with all other
Geratherm® Respiratory GmbH products via the universal
software BLUE CHERRY®.
(See chap. 4.1.4 “System Construction and Electrical Safety”)
In all cases, the Diffustik resp. the variant PFTstik is only
intended for use in closed, pleasantly temperature-controlled
(19 °C - 25 °C) rooms in a clinical area.
In the respective room, there must be neither flammable or
explosive gases, nor magnetic fields (e.g. MRI).
With the Diffustik variant or the PFTstik variant in combination
with the Add-on CO Diffusion extension, the room must always
be well ventilated, to protect users and third parties from
continuous exposure to harmful carbon monoxide (CO).
The following applies in general: The installation of Diffustik resp.
the variant PFTstik is only deemed safe and in line with intended
use when this is carried out in accordance with the details in
chap. 12.2 “Installation and Operating Conditions”.
Attention
The Diffustik resp. the variant PFTstik must be operated exclusively
with the equipment cart approved by Geratherm® Respiratory (REF
10.905). The equipment cart provides electrical isolation from the
power supply network. All necessary peripheral devices such as
computer, printer and monitor may only be connected to the power
supply network via the equipment cart. See also. chap. 4.1.4 "System
Construction and Electrical Safety". If you have any questions, please
contact your local specialist dealer.