Malfunction
Reason
What has to be done
There are no
indications on the
display or they are not
stable
(in APEX or PULP
mode)
Electrical circuit
continuity of the device
is broken
Check the electrical circuit
continuity
(see p.7.6);
if
necessary, replace the
cable and/or file clip
The device does not
react at the control unit
buttons pressing
The program is hung.
Restart the program:
- Press “Reset”
(9-fig.2)
button using any thin
subject (e.g. a needle).
Herewith the device power
switches off automatically.
- Switch on the device
power
(see p.7.5)