22
USER’S GUIDE
www.geosoft.ru/en
3. When attaining the canal apical constriction by the file tip (Apex =
0.0), disconnect the file clip from the working file.
7.8. Operating in the
EOD
mode
1. Switch on the device
EOD
mode
(see p.7.5)
2. Place the passive electrode (lip clip
(5-fig.1)
)
on patient’s lip, and
touch the sensitive point of the studied tooth by the tip of active
electrode (EOD probe
(4– fig.1)
)
After placing the electrodes, be sure to warn patient of pain reaction
and ask him/her to give a sign when first feeling pain. Now you can
start diagnosing.
3. Press (or press and hold) the touch button
«START/STOP»
(5–
fig.2).
The device will start gradually increasing the voltage and displaying
the running value of “diagnostic” current (in mcA) passing through
tooth
(see fig.19).
The measurements will be followed with discontinuous sound signal
and light indication.
Fig.19