4. When attaining patient’s pain reaction (as soon as he/she gives a
sign), press (or release) the button
«START/STOP»
and take away the
active electrode of studied tooth.
Herewith the “diagnostic” current measured value will be fixed on the
display.
Interpretation of results of measurements is presented in table 2
Table 2
Attention!
The diagnosis presented in table 2 is based on average
data and can serve only as a reference point for the dentist. Dentist
takes final decision of the using data of the anamnesis, survey and
inspection of the patient by means of additional methods of
diagnostics.
The reset of measurements result happens automatically later ~6-7
sec.
- For compulsory reset of result of measurements press the touch
button
"START/STOP"
(5-fig. 2).
Current value,
mcA
Diagnosis
Color of the indicator
(3 - fig.2)
2-8 Intact
tooth
GREEN
9-14 Caries
lesions
ORANGE
15-24
Deep caries lesions
25-44 Pulpitis
RED
45-80 /
no reaction
Periodontitis