MEMS Digital Portable Horizontal Inclinometer
09/02/2022
61
12.2. ASD Readout
Table 7: ASD Readout Troubleshooting
Fault
Possible Cause
Remedy
ASD will not turn on
Flat battery
Charge battery
Faulty charger
Try new charger and cable /
check outlet is powered
Faulty battery
Replace battery
The ASD cannot find the
reel or finds but cannot
connect
Reel not turned on
Power the reel on
Bluetooth is not
enabled
Enable Bluetooth on the ASD
Incorrect connection
sequence
ALWAYS use the IncloPRO
app to initiate the Bluetooth
connection, NOT the phone
Bluetooth options
Interference from other
paired Bluetooth
devices
Restart reel and ASD and
power off other Bluetooth
items such as speakers that
are linked to the same ASD
Bluetooth fails to
connect despite using
the app
Return to Geosense for
inspection
The ASD freezes after
extended periods of use
Bluetooth connection
Power cycle the reel and
reconnect using the app to
continue monitoring
Power saving settings Make sure there are no
power saving options
enabled in the ASD settings
Fluctuating readings
Faulty signal from
prove
Check probe connection
Check cable for damage
Check Bluetooth connection
Check ASD settings