32
F.4 Troubleshooting
Below are some commonly experienced problems along with possible remedial action. Contact
the factory if any problem remains unresolved or additional help is required.
Symptom: A particular channel on the multiplexer appears to be malfunctioning:
Check sensor connections on the terminal board. Clean if corrosion exists.
Try moving the sensor wired to the suspect channel to another channel to verify the
malfunctioning of the channel (as opposed to the sensor).
Symptom: No channels are working:
Inspect circuit board for shorts, opens, or other damage.
Is moisture present on circuit board? If so, install desiccant to absorb.
Symptom: Channel selection appears to be random:
Has corrosion built up on the circuit board? Clean if necessary.
Is there a source of electrical noise nearby? Move multiplexer or noise source if possible.
F.5 8032 Multiplexer Specifications
GENERAL:
Power Requirements
10-16 VDC (unregulated)
Quiescent Current
(MICRO-800/MICRO-1000 mode)
80
µ
A (16CH mode)
130
µ
A (32CH mode)
Quiescent Current (GK-403 mode)
12 mA
Channel Activated Current
30 mA
Control Line Input Impedance
100 k
Ω
(CLOCK), 100 k
Ω
(RESET)
Control Line Input Levels
TTL or RS-232 (
±
9 VDC)
Transient Protection
16 VDC Transzorbs
Operating Temperature
−
40 to
+
60
°
C
RELAYS:
Type
NAIS TXS2SA-4.5V DPDT non-latching
Power
11.1 mA @ 5VDC (55.5 mW)
Contact Type
Gold clad silver alloy
On Resistance
100 m
Ω
Coil Resistance
405
Ω
Maximum Switching Power
30W (resistive)
Maximum Switching Voltage
110 VDC
Maximum Switching Current
1 A
Operate Time
≈