Genevac HT-24
User Manual
44
10-1562 Issue 3-7 March
2014
Example 3. Mixture of High and Low Boiling Point Solvents
Method consists of three runs linked together to form a multi stage run, with settings as
follows:
Stage 1
Remove Organic
Phase
Stage 2
Dry High BP
Solvent
Sage 3
Final Drying
Run No.
1
2
3
Run Data Protection
N
N
N
Next Run Number (0 for none)
2
3
0
Run Name
High - Low Mix
High - Low Mix 2
High - Low Mix 3
Sample Holder Type
Your Holder
Your Holder
Your Holder
Sample Volume (ml)
Your Volume
Your Volume
Your Volume
Solvent
DMSO or DMF & DCM DMSO or DMF & DCM DMSO or DMF & DCM
SampleGuard Control Temp (°C)
40
40
40
SampleGuard Control Channel
1
1
1
CoolHeat Enable Pressure (mbar)
50
50
50
Chamber Temperature control (°c)
preheat to 15
preheat to 40
preheat to 30
Rotor Speed
High
High
Low
Heat-Off Option
Elapsed Time
Heat Flow
Elapsed Time
Heat-Off Elapsed Time (HH:MM)
01:00 to 1:30
-
00:00
End-of-Run Option
Elapsed Time
-
Elapsed Time
End-of-Run Elapsed Time (MM:MM)
01:00 to 01:30
-
01:00 to 03:00
Auto Defrost & Drain
Y
N
Y
Pressure Control Regime
Variable Dri-Pure
Full Vacuum
Full Vacuum
Dri-Pure Start Pressure (mbar)
750
-
-
Dri-Pure End Pressure (mbar)
70
-
-
Dri-Pure Ramp Time (MM)_
55
-
-
Final Control Pressure (mbar)
65
-
-
Summary of Contents for HT-24