User Manual
Genevac HT-24
10-1562 Issue 3-7 March
2014
35
Note:
Some Bohdan Miniblock system sample holders exceed the mass limitations of
Genevac systems. If you plan to use Bohdan Miniblocks in a Genevac system, please
contact Genevac for a list of the relevant weights.
Caution:
With the exception of microtitre plates, do not load any non-Genevac
holders into a system without gaining approval from Genevac.
Mass limit
The mass limit for Genevac HT-SII evaporators is 1.5 kg per swing. This is the total mass
that can be loaded onto each position of a swung rotor, and includes:
The swing
The sample holder
The tubes or vials
The sample solutions.
Do not exceed the mass limit under any circumstances. In most cases, with normal solvent
volumes, loaded Genevac sample holders will fall within this limit. If in any doubt, load up a
swing and weigh it.
Sample holders and swings
When using separate sample
holders and swings
make sure:
The sample holders sit
centrally and squarely in the
swing
The sample holders do not rest
on the edge of the swing.
Caution:
Miss-loading may result in unrecoverable sample loss, damage to
equipment, and could void the warranty.
Summary of Contents for HT-24
Page 2: ......