12
3.6 After the cleaning process
After every cleaning process the tank should be cleaned. On the ground of the tank mature dirty
leftovers reduce the ultrasonic achievement and can lead to damages on the tank ground.
Cleaning liquids can be used according to soiling several times, but they cannot be refreshed by
recent addition of cleaner again. Spent cleaning liquids have a lower cleaning effect and have to
exchange.
3.7 Preparation of medical instruments
o
If use the ultrasonic - cleaner in connection with disinfection preparations (medicine
product class IIa) in addition to support the disinfection effect or to accelerate as
accessories to a medicine product and is to be arranged even as a medicine product
class I.
o
In case of cleaning or disinfection of medical instruments, the hygienic security is to be
guaranteed after the utilisation. The cleaning and disinfection has to be carried out by the
operator, according to the hygiene guideline (plan) under using a e.g. VAH/DGHM
certified surface disinfectant.
4. How to clean the device
•
Before cleaning every device disconnect from power supply system.
•
By devices without draining cock pour out the tank contents over the corner (backside left),
besides, splashes avoid to the device ground or the cable connection.
•
Devices do not wash up or dip into water.
•
Rinse the tank thoroughly and rub dry, no steel wool, scratches, scrapers use or similar.
•
Edges and remains in the tank are to be removed with a customary high-grade steel cleaner
without abrasive addition.
Tip:
Remaining metal parts on the stainless steel surface as well
as rust particle from the water pipe system or the cleaning property can penetrate the
passive protective coating of the stainless steel, "activate" the stainless steel and it starts to
rust. This foreign rust causes pitted areas and the tank becomes leaking.
•
Wipe case outside to the cleaning only humid, use when required a customary high-grade
steel cleaner without abrasive addition.
5. Functional disorder
problem
possible cause
remedy
weak uneven insonification,
noise is too loudly or uneven
liquid is not degassed
10 min. insonification
too much cleaning parts
take some parts out
uneven noise
change filling state
Liquid does not become warm
filling state to low
fill up liquid
heating damaged
customer service
pilot lamp does not shine
lamp or heating damaged
customer service
liquid to hot
warm by cavitations
put down temperature
tank covered
remove lid
Device cannot be switched on
electricity supply discontinuous check, produce electricity
supply
no insonification
electronic problem
customer service